PEPconnect

Inspecting IMT Peristaltic Pump Tubing Video

This video provides instructions on how to inspect the IMT peristaltic pump on the Dimension Vista® System.

To begin inspecting the IMT peristaltic pump tubing release the left interlock, an lift the left lid. Locate the IMG peristaltic pump near the sample loading area. Lift the pump lid and hold the tubing between your fingers to inspect the tubing between the collars for flatten sections. If the tubing is flat and it will need to be replaced following the replace IMT peristaltic pump tubing procedure. If tubing does not need to be replaced, reinstall the tubing in the pump and close the pump lid. When finished, close the left lid.

Dimension [email protected] System Inspect IMT Peristaltic Pump Tubing IMT