Cleaning HM Wash Probe Video

This video provides instructions on how to stylette HM wash probes on the Dimension® Xpand® System.

The HM Wash probes must be cleaned weekly to prevent residue buildup. From the operating menu, go to the pump prime screen to prevent any automatic priming of the HM Wash probes. To display the pump prime screen, press F for system prep. F7 pump prime. F6, HM. Raise the IMT lid. Insert a strong, slender tool into the reagent lid. Latch and press firmly to raise the lid. Next, dip a clean cotton swab in water. Beginning at the top of the probe, wipe down the outside of a wash station probe. Dispose the cotton swab into a biohazard waste container. Use a new cotton swab to clean each of the HM Wash probes. Press F1 HM Wash pump to prime the HM Wash pumps. Complete the procedure by documenting the cleaning on the weekly maintenance log sheet.

SIEMENS F4. SYS REAGENT F3: IMT E PREP Clean HM F2: REAG CLEANER HMWASH PUMP PUMP ASH PUMP WATER F4: SYSTEM PREPK E PREP .ESULTS Wash Probes F8: DAI F7: PUMP PRIME PUMP F8: SEGMENT STATUS OSTICS Dimension XpandIXpand Plus Dimension Plus

  • weekly
  • maintenance