Edit Dose Information for Completed Acquisition
This Step-by Step will instruct the user how to edit dose information for a completed PET acquisition. This can be used for editing all injection information including isotope, pharmaceutical, injected amount and injection time.
Edit Dose Information for Completed Acquisition Biograph Vision, mCT and Horizon This Step-by Step will instruct the user how to edit dose information for a completed PET acquisition. This can be used for editing all injection information including isotope, pharmaceutical, injected amount and injection time. 1. If the exam has been closed, reload it into the Examination task card but double clicking on the study folder from the patient browser. 2. Click the Edit Dose button on the PET Routine subtask card. Melanoma_Rt_eye Melanoma_Rt_eye Anonymous Biograph64 Anonymous Biograph64_mCT +01-Jan-1946, F VG70B *01-Jan-1946, F VG70B 09-Aug-2019 F-SP 09-Aug-2019 F-SP-CR 11:48:07.08 12:06:08.00 1 IMA 1 5 IMA 435 TOP 1 AC 2001 SP-1875.0 SP -1851.0 ER 10cm KV 120 mA 35 TI 18.6 GT 0.0 2048 0/0 T801U32P0 1d Melanoma_NC_CBM (Adult) Melanoma_Rt_eye 1737.5 mm 987 sec 10 mCi Topogram Isotope F-18 Pharm. Fluor oxyglucose 4mm CT Dose 10.00 mc Pause Date 09-Aug-2019 Time 10:38:56 PET WB Scan range Match CT Range Scan duration 00:16:27 Range Distance Duration Speed 1 809 00:05:59 12 928.5 00:10:28 1.5 Range: Begin End Table: Position Height Scan direction Hold Recon -113.5 -1851.0 -400.0 100.0 Caudocranial Recori Routine Scan Recon Auto Tasking Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/19/2020 | HOOD/HILS HOOD5062003089263 SIEMENS Healthineers Job Aid | Edit Dose Information for Completed Acquisition Effective Date: 05/19/2020 | HOOD/HILS HOOD05162003089263 The Radiopharmaceutical Dose dialog displays: Radiopharmaceutical Dose X Isotope -18 Pharmaceutical Fluorodeoxyglucose Volume (mL) 0.00 Pre-injection Assay Dose 10.00 mCi Date 09-Aug-2019 Time 10:38:56 Injection Assay Dose 10,00 mCi Start Date 09-Aug-2019 Start Time 10:38:56 Stop Date 09-Aug-2019 Stop Time 10:38:56 Post-injection Residual Assay Dose mCi Date 27-Apr-2020 Time 20:34:52 DICOM Information Edit Blood Glucose Level Radionuclide Total Dose 10.00 mCi 0.00 mmol/L Radiopharmaceutical Start Time 10:38:56 Radiopharmaceutical Injection Duration 0 S Date 09-Aug-2019 OK Cancel 3. For all mCT and Vision systems VG70 and above, click the radio box next to Edit in the DICOM information section: Post-injection Residual Assay Dose mCi Dat Time DICOM Information Edit Blood Glucose Level Radionuclide Total Dose 10.00 mc 0.00 Radiopharmaceutical Start Time mmol/L 10:38:56 Radiopharmaceutical Injection Duration 0 S Date 09-Aug-2019 OK Cancel Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date 05/19/2020 | HOOD/HILS HOOD5062003089263 Healthineers Page 2 of 6 Job Aid | Edit Dose Information for Completed Acquisition Effective Date: 05/19/2020 | HOOD/HILS HOOD05162003089263 4. This will turn the attributes below Edit to active. Once they are active, edit the needed information and click OK. The system will inform you if new reconstructions are needed. (For most radiopharmaceuticals, DICOM information of all previously com- pleted reconstructions is updated and additional reconstructions are not necessary). A Dialogue box will display directing the user to simply recalculate SUVs: PET Examination X Dose parameter changes are updated in PET Raw Data and PET Image series. Please recalculate SUV. 10 oma_NC_CBM (Adult) 1737.5 mm 9 ogram OK m CT Dose 10.00 use Date 09-Aug-2019 Time 10:38:56 Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date 05/19/2020 | HOOD/HILS HOOD5062003089263 Healthineers Page 3 of 6 Job Aid | Edit Dose Information for Completed Acquisition Effective Date: 05/19/2020 | HOOD/HILS HOOD05162003089263 *For Horizon systems and all mCTs on VG62 or below, there will not be an option for editing DICOM information section in the Dose Dialogue window: Radiopharmaceutical Dose X Isotope F-18 Pharmaceutical Fluorodeoxyglucose Volume (mL) 0.00 Pre-injection Assay Dose 10.00 mCi Date 23-Oct-2012 Time 15:01:36 Injection Assay Dose 10.00 mc Date 23-Oct-2012 Time 15:01:36 Post-injection Residual Assay Dose mCi Date 23-Oct-2012 Time 15:01:57 DICOM Information Blood Glucose Radionuclide Total Dose 10.00 mCi Radiopharmaceutical Start Time 15:01:36 0.00 mmol/L Date 23-Oct-2012 OK Cancel Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date 05/19/2020 | HOOD/HILS HOOD5062003089263 Healthineers Page 4 of 6 Job Aid | Edit Dose Information for Completed Acquisition Effective Date: 05/19/2020 | HOOD/HILS HOOD05162003089263 The user should edit dose information and submit new reconstructions when prompted: 10cm PET Examination X PET Raw data information has changed. Please reprocess the data and recalculate SUV. VIABILITY (Adul ymous OK Dose 11.90 mCi Date 31-Jan-2017 Time 11:40:00 *All clincal images courtesy of Siemens Healthineers Customer Care Center Please note that the learning material is for training purposes only! Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date 05/19/2020 | HOOD/HILS HOOD5062003089263 Healthineers Page 5 of 6 Job Aid | Edit Dose Information for Completed Acquisition Effective Date: 05/19/2020 | HOOD/HILS HOOD05162003089263 For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date 05/19/2020 | HOOD/HILS HOOD5062003089263 Healthineers Page 6 of 6
- Edit Dose Information