PEPconnect

Elara Max: Common Error Codes

This job aid will cover common error codes that may be seen on the Elara Max
system.

Elara Max: Common Error Codes XPRF Error Message Cause Measure 194 Overheating of capacitor charging unit Switch off the system and wait for it to cool down. 183 The tube housing is too hot. Switch off the system and wait for the single tank to cool down. 205 The anode rotates for more than 15 s Make sure that the preparation time without radiation being released of 15 s is not exceeded. 209 Battery voltage too low, no exposures Charge the batteries. Connect the sys- possible tem to the mains supply. 221 The exposure release button was re- The image may be invalid. Check the Leased before the exposure was com- image and repeat the exposure, if nec- pleted essary 225 The system battery is completely dis- If you need to make an exposure im- charged, the battery bank is not con- mediately, plug system to mains pow- nected or the fuse of the battery was er, confirm error message 225 at blown startup by pressing the light field but- ton on the control panel. Then pro- ceed normally 239 D850 Self test error Please call service. SIEMENS siemens.com/healthineers Healthineers Job Aid | Common Error Codes Effective Date: 07/26/2019 | Quality HOOD05162003027492 Error Message Cause Measure 240 D850 temperature sensor out of range If it is a sporadic error, please switch the entire system OFF and ON. If the error appears again, please call ser- vice. 241 Error in voltage measurement circuit If it is a sporadic error, please switch the entire system OFF and ON. If the error appears again, please call ser- vice. 242 D850 Fuse blown Please call service. 243 Short circuit at the output Please call service. If errors/warnings with different codes are displayed, switch the system off and on again. If the message continues to be displayed, notify Service. Siemens Healthineers Services Page 2 of 3 © Siemens Healthcare GmbH, 2017 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2017

  • Elara Max: Common Error Codes