epoc® Blood Analysis System Care and Maintenance Job Aid
This job aid provides instructions on how to perform basic maintenance and cleaning procedures on the epoc® Blood Analysis System.
Care and Maintenance Job Aid epoc® Blood Analysis System The epoc® Blood Analysis System requires a minimum of care and maintenance. 1. The following general practices are recommended: A. Turn off the epoc Reader and epoc Host when not in use to conserve battery life. B. Store the epoc Reader and epoc Host in a secure location when not in use. C. Keep the epoc Reader and epoc Host in a dry location while using and during storage. D. Periodically check the condition of the wires and cables of the AC adaptor to check for wear and to ensure integrity of electrical conditions. E. Refer to the epoc System Manual regularly and ensure that the most current and up-to-date version is being utilized. 2.Cleaning/Storage Instructions: A. Do not expose any electrical contacts to cleaning solutions. B. Never submerge any part of the epoc Reader and epoc Host into any liquid.Never allow liq- uids to pool in Pivot or Membrane Switch areas. C. Keep the epoc Reader and epoc Host in a dry location while using and during storage. D. Never apply liquid anywhere onto the internal or external areas of the epoc Reader or epoc Host. E. Wipe with a damp, soft cloth or gauze pad to clean the surface of the epoc. F. Use a mild detergent, alcohol, 10% household bleach solution or non abrasive cleaner when cleaning the surface of the epoc Reader or Host. B spar test 3:24 O CEEEEL De epoca EEEEI EEEE Oc B G Effective Date: 08/01/2018 | HOOD05162002912329 SIEMENS siemens.com/healthineers Healthineers Job Aid | epoc Care and Maintenance Effective Date: 08/01/2018 | HOOD05162002912329 3.Never attempt to: A. Clean the inside of the Readers card insertion slot. B. Clean a test card. C. Sterilize or autoclave any part of the epoc system. epoc BLOOD ANALYSIS EM B Crea Het 0000 2016-08-28 A SN 708702 .&@ CE Epocal, Inc. Ottawa, ON Canada www.epocal.com 4.Decontaminating the epoc Reader or Host when blood is spilled: A. Prepare a 70% isopropyl alcohol solution. It is recommended that the solution be prepared fresh daily. B. Soak several gauze pads in the bleach solution. When removing a pad from the solution, squeeze excess liquid, so that the pad does not drip on the system. C. Gently rub any areas of dried blood with one or more moist pads until they are soft enough to wipe clean. D. After removing any stained areas, clean all surfaces twice with fresh pads soaked in the bleach solution. Ensure that the bleach solution is in contact with the surface for three minutes before wiping off. E. Soak Fresh pads in warm tap water and squeeze excess fluid. Wipe all surfaces and allow them to dry before turning any of the epoc system components on. F. Dispose of the used gauze pads in a biohazard disposal receptacle. Siemens Healthineers Services Page 2 of 5 Job Aid | epoc Care and Maintenance Effective Date: 08/01/2018 | HOOD05162002912329 5.Changing the epoc Reader Battery: A. Remove the screw that connects the battery door to the Reader using a Phillips #0 or 3/16” slotted screwdriver (the type of screwdriver needed is dependent on the type of screw found in the Reader). B. Lift the battery door from the Reader. C. Carefully remove the battery from its compartment to allow access to its cable connection to the Reader. D. Gently unplug the battery from the Reader. Tweezers may be useful in order to pull the con- nector out. E. Connect the new battery connector to the epoc Reader. The connector orientation tab should be visible as seen in the picture. F. Place the new battery into its compartment. Place its wires into the compartment to the left of the battery. Ensure that the battery is lying flat without wires underneath or on top. G. Install the battery door by inserting its hinge at an approximate 30° angle behind the bat- tery. Pressure needs to be applied to the edge of the door in order for its hinges to snap into place. H. Secure the door by fastening with the screw using a Philips #0 or 3/16” slotted screwdriver. Do not overtighten the screw to avoid damaging the plastic. E D F Siemens Healthineers Services Page 3 of 5 Job Aid | epoc Care and Maintenance Effective Date: 08/01/2018 | HOOD05162002912329 6.Replacing the epoc Reader battery door: A. Remove the screw that connects the battery door to the Reader using a Phillips #0 or 3/16” slotted screwdriver (the type of screwdriver needed is dependent on the type of screw found in the Reader). B. Lift the battery door from the Reader. C. Do not move the battery or wiring; it is installed in a way that facilitates easy battery door replacement. D. Install the new battery door by inserting its hinge at an approximate 30° angle behind the battery. Pressure needs to be applied to the edge of the door in order for its hinges to snap into place. The battery door can then be closed into place. E. Secure the door by fastening with the screw using a Philips #0 or 3/16” slotted screwdriver. Do not overtighten the screw to avoid damaging the plastic. D C Siemens Healthineers Services Page 4 of 5 Job Aid | epoc Care and Maintenance Effective Date: 08/01/2018 | HOOD05162002912329 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (herein- after collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2018 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018
- epoc
- care
- maintenance