epoc® Blood Analysis System with NXS Host Changing The epoc Reader Battery Job Aid
This job aid will show you the steps on how to properly replace the battery on the epoc Reader.
epoc® with NXS Host Blood Analysis System SIEMENS Healthineers Changing the epoc Reader Battery Job Aid Changing the epoc Reader Battery 1. Remove the screw that connects the battery door to the Reader using a Philips #0 or 3/16” slotted screwdriver 2. Lift the battery door from the reader 3. Carefully remove the battery from its compartment to allow access to its cable connection to the Reader 4. Gently unplug the battery from the Reader. Tweezers may be useful in order to pull the connector out 5. Connect the new battery connector to the epoc Reader. The connector orientation tab should be visible as seen in the picture 6. Place the new battery into its compartment. Place its wires in the small area to the right of the connection to protect the wires. Ensure that the battery is lying flat without wires underneath or on top * Procedure is continued on next slide Template Effective Date: 27 May 2019| HOOD05162003010284 1 Effective Date: 07-JUN-2021| HOOD05162003194856 Unrestricted © Siemens Healthineers, 2021 epoc® with NXS Host Blood Analysis System SIEMENS Healthineers Changing the epoc Reader Battery Job Aid Changing the epoc Reader Battery 7. Install the battery door by inserting its hinge at an approximate 30˚ angle behind the battery. Pressure needs to be applied to the edge of the door in order for its hinges to snap into place. The battery door can then be closed into place 8. Secure the door by fastening with the screw removed, using a Philips #0 or 3/16” slotted screwdriver. Do not over- tighten the screw to avoid damaging the plastic Template Effective Date: 27 May 2019| HOOD051620030102842 Effective Date: 07-JUN-2021| HOOD05162003194856 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Template Effective Date: 27 May 2019| HOOD05162003010284 3 Effective Date: 07-JUN-2021| HOOD05162003194856 Unrestricted © Siemens Healthineers, 2021
- epoc
- battery
- maintenance
- reader