FLC MAX Systems: Exporting Images
FLC MAX Systems: Exporting Images step by step
FLC MAX Systems: Exporting Images Radiography 1. Insert CD/DVD or USB. SIEMENS Fluorospat Compact 2. Select the “Patient” tab on the right; Examined Patients Archiving Information Patient select the “Examined Patients” tab at Birth Sex Study Description Ward Physician Flag Date anc+ :01 ... LONG LENGT. 25-05-2016 the top of the FLC. M VP SHUNT 4N 23-03-2013 M HIP ARTHROG ... ETS 22-03-2013 M HIP ARTHROG ... RA ... -D 20-03-2013 M VP SHUNT 4E D 16-03-2013 Examination M Knee 3 views .D 18-06-2012 95 ... M L-spine 2 views ... --- D 13-06-2012 Patient Settings Service 97 ... F Chest 2 views -D 12-06-2012 ? 97 ... F Pelvis ap -D 12-06-2012 F Barium Swallo ... -D 03-05-2012 M L-spine 4 views .. -D 30-04-2012 F Barium Swallo ... -- D 13-04-2012 Postprocessing go ... F Flouro guidance ... --- D --- 11-04-2012 EXI F Barium Swallow --- D --- 11-04-2012 METASTATIC CH ... 03-04-2012 F SHOULDER.LE ... LAB LAB 27-03-2012 M FOOT,LEFT RAD 27-03-2012 F Knee 2 views RAD 13-03-2012 3. Select the Examined Patient to be Preregistered Patients Examined Patients Archiving Information exported. Patient Name Patient ID Date of Birth Sex Study Description Ward Physician Flag Date anc Emergency ,... E_30 25-05-201 ... LONG LENGT ... --- D ---- 25-05-2016 Luminos Agi ... 0000 01-01-200 ... M VP SHUNT 4N --- D ---- 23-03-2013 Luminos Agi ... 0000 01-01-200 ... M HIP ARTHROG ... ETS --- D -... 22-03-2013 Luminos Agi ... 0000 01-01-200 ... M HIP ARTHROG ... RA ... --- D 20-03-2013 Luminos Agi ... 0000 01-01-200 ... M VP SHUNT 4E 16-03-2013 Luminos Agi ... 050961 08-07-195. M Knee 3 views 18-06-2012 19 AA AAAA 4. Select the Export icon, bottom right of the FLC. Export >REF X Select organ program |2/3/ 0% h SIEMENS siemens.com/healthineers Healthineers Job Aid | FLC MAX Systems: Exporting Images Effective Date: 06/14/2017 | Quality HOOD05162002609388 5. Select the appropriate options and Export X press “Export” on the pop-up box. Target Type USB device Target Export Format DICOM Processed Anonymized Downsized Export Dose Report as: DICOM dose report DICOM SC image Number of Copies 1 Transfer All Space: Used: 66% Required 1% Set as Default Export Status Close Format – DICOM option is mostly used; this option is for loading images to PACs or viewing them on a computer. Processed – shutter, panning and graphics will be “burned” into the pixel data. Anonymized - Patient data will not be included with the images. NOTES Downsized – determines the matrix size a fluoro loop will be stored (Original or Downsized). Click the DICOM SC image check box to include the dose report. Transfer – Select “All” to send all images in the study, select “Marked” for only the images that are selected to send (white dot on the image). Siemens Healthineers Services Page 2 of 3 © Siemens Healthcare GmbH, 2017 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2017
- FLC Exporting Image on MAX Systems