FLC PEX Editor: Adding or Removing Alternate Acquisition Programs

This Step by Step will guide users on adding or removing Alternate Acquisition
Programs with in the PEX Editor.

FLC PEX Editor: Adding or Removing Alternate Acquisition Programs XPRF To add or remove an “Alternate Acquisition Program in the “PEX Editor”, follow the below steps: 1. From the “Patient” tab, click on the “Settings” subtask card, and then select the “PEX Editor” icon and log in via “Siemens”. Patient Settings Service Till EXI Adding an Alternate 2. Double click on the desired “OGP View” name to open the “Organ Program” tab. Eram Organ Program Fluoro Program Param. Module Configuration OGP View Examination View X KNEE Left 4-12yrs (d) WP TOMO S(O) X KNEE MERCHANTS VIEW (d) JOINT SURVEY_S(d) X KNEE OBL (d) KNEE 2 VIEWS BILATERAL_S(d) X KNEE OBL Left (d) KNEE 2 VIEWS LEFT_S(d) X KNEE OBL Right (d) Insert OGP KNEE 2 VIEWS RIGHT_S(d) X KNEE Right 0-3yrs (d) KNEE 3 OR MORE VIEWS BILATERAL X KNEE Right 4-12yrs (d) KNEE 3 OR MORE VIEWS LEFT_S(d) TATITT X KNEE SUNRISE (d) KNEE 3 OR MORE VIEWS RIGHT_S(d) X KNEE SUNRISE BILAT (d) KNEE ARTHROGRAM_S(d) X KNEE SUNRISE Left (d) LONG LEG EXAMINATIONS_S(d) X KNEE SUNRISE Right (a) LOOPOGRAM_S(d) X KNEE TUNNEL (0) LOWER EXTREMITY ARTERIOGRAM X KNEE TUNNEL LEFT (0) Remove OGP LOWER EXTREMITY VENOGRAM_S(d X KNEE TUNNEL RIGHT (0) LUMBAR PUNCTURE_S(d) X KNEES AP BILAT (d) MANDIBLE S(d) ................... Group by @ Name Body Part Acq type Sort by @ Name @Body Part Acq type 3. Under “Alternative Acquisition Programs” select the first empty box available. Exam Organ Program Fluoro Program Param. Module Configuration Body Part Examined KNEE Alternative Acquisition Programs Program Name KNEE MERCHANTS VIEW (d) . Acquisition Mode DR RAD Acquisition Type Free Exposure SIEMENS siemens.com/healthineers Healthineers Job Aid | Adding/Removing Alternate Acquisition Programs Effective Date: 09/04/2018 | Quality HOOD05162002945620 4. Choose the desired DR “Alternate Acquisition” location (free, wall, table as indicated below). Alternative Acquisition Programs Alternative Acquisition Programs Alternative Acquisition Programs cessing Parame ettings essing Parame ettings cessing Parame ing Standard PT Position ing Standard PT F bing Standard size as cropped Laterality ize as cropped size as cropped tion 2.0 La DR Rad Free Exposure pr lon 2.0 View [ tion 2.0 Parameters Anatomic Parameters DR Rad Wall Ar Parameters y/Skull #/Skull Invert (Im y/Skull DR Rad Table Invert ( invart 5. Select the drop down arrow and scroll through the list to locate/select the desired “OGP View”. Exam Organ Program Fluoro Program Param, Module Configuration Body Part Examined KNEE Alternative Acquisition Programs Program Name KNEE MERCHANTS VIEW (d) W KNEE MERCHANTS Acquisition Mode DR RAD W KNEE AP Right (d) Acquisition Type Free Exposure W KNEE LAT (d) W KNEE LAT Left (d) X - Ray Parameters Image Proce W KNEE LAT Right (d) Mark Image Auto-cropping W KNEE MERCHANTS VIEW (d) Fluoro Program Set Not defined none W KNEE OBL (d) Defined sizew KNEE OBL Left (d) MFlek ferality Not Dentes AmplificationW KNEE OBL Right (d) osition Not Defined Note: If the desired “OGP View” cannot be found, refer to Step by Step for “Creating a new OGP View”. 6. Select “Store” to save the changes. Anatomic order number Store Store as new Print Delete previous next o Close Removing Alternate View 1. To remove an Alternate Acquisition Program, select the drop down and scroll to the top of list and select the blank space. Alternative Acquisition Programs IQ TESTPROGRAMS P16_AP150_RAD_FE (d) P16_AP150_RAD_T (d) P16_AP150_RAD_T_EXI1 (d) P16_AP150_RAD_T_EXI2 (d) P16_AP150_RAD_UT (d) P16_AP150_RAD_UT_EXI1 (d) 2. Select “Store” to save changes. Anatomic order number 0 Store Store as new Print Delete previous next Close Siemens Healthineers Services Page 2 of 3 © Siemens Healthcare GmbH, 2017 Job Aid | Adding/Removing Alternate Acquisition Programs Effective Date: 09/04/2018 | Quality HOOD05162002945620 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 3 of 3 © Siemens Healthcare GmbH, 2017