FLC: Table Integrated Ortho for Agile, Ysio, dRF and RAX
This job aid will guide the end user through the set up for Ortho Long Leg/Spine Imaging at the table.
VC10 System: Table Integrated Ortho Radiography 1. Open a patient exam and select an OT_organ program Had in the Examination Mode. 2. The system can acquire up to 4 images per study. 34.1 cm x 19.0 cm 300 cm Assess the technical factors as needed for each Ortho none 300 cm Step. Ortho Step I TODicen 8.0 MAX Dose 0.0 000 Foon % 3. Align the tube to the table center position. Press the OGP move button to align the tube for fully- automated systems, as shown. Position patient on the table. Coverage of anatomy depends on maximum SID and can vary. 4. Angle the x-ray tube and use the laser to set start (press position1) and end (press position2) on the TODA 10.5 cm Device Position patient. Check marks on the Ortho Tab indicate start and end positions are achievable and are set. Position 1 Ortho Acquisitions are head-to-foot by default. TOD (table- object-distance) can be entered by pulling the tape Position 2 measure from the tube down and stop at the level of the anatomy (leg or spine). TOD should be entered for each exam to ensure accuracy of distance measurements on images. SIEMENS siemens.com/healthineers Healthineers Job Aid | VC10 System: Table Integrated Ortho Effective Date: 8/23/2016 | 297 5. Press and hold the acquisition button. Siemens Healthineers Services Page 2 of 3 © Siemens Healthcare GmbH, 2017 Job Aid | VC10 System: Table Integrated Ortho Effective Date: 8/23/2016 | 297 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 3 of 3 © Siemens Healthcare GmbH, 2017
- Ysio