FLC: Wall Integrated Ortho for Agile, Ysio, dRF and RAX
This Job Aid will guide the user through setting up the system for Ortho Long
Leg/Spine Imaging at the Wall Bucky.
VC10 System: Wall Integrated Ortho Radiography 1. Open a patient exam and select an OW_organ Fluoro program in the Examination Mode. The system can acquire up to 4 images per study. Assess 34.1 cm x 19.0 cm 300 cm the technical factors as needed for each Ortho Target none @ 300 cm Step. Ortho Step 13 TODtcm) 8.0 RV 73 me Dos mm Cu 0.0 On0 Focus 2. Align the tube to the wallstand at the appropriate SID (180cm or 300cm depending on room size). Press the OGP move button to align the tube for fully-automated systems, as shown. Position patient on the Ortho Stand provided by Siemens. 3. Angle the x-ray tube up/down, using the laser to set start (press position1) and end (press TODA 10.5 cm Device Position position2) on the patient. Check marks on the Ortho Tab indicate start and end positions are Position 1 Ortho set. Acquisitions are head- to-foot by default. *TOD (table-object- distance) is a measurement in cm from the Ortho Stand to Position 2 the anatomy (spine or leg). It should be entered for each exam to ensure accuracy of distance measurements on images. 4. Press and hold the acquisition button. SIEMENS siemens.com/healthineers Healthineers . Job Aid | VC10 System: Wall Integrated Ortho Effective Date: 8/18/2016 | 394 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 2 of 2 © Siemens Healthcare GmbH, 2017
- Rax