How to Perform a New Biopsy Unit Referencing - Error 7
This job aid covers performing a new biopsy unit, referencing Error 7.
How to Perform a New Biopsy Unit Referencing- Error 7 MAMMOMAT Revelation If the biopsy unit is mounted on the MAMMOMAT, the needle holder automatically moves through the reference positions. The referencing movement of the needle holder must not be interrupted. If a malfunction occurs during the referencing movement, the following window appears: WIT AWS The Biopsy Unit encountered a hardware problem during self test. Error: 7. A new Biopsy Unit referencing is required. Please remove the needle (respectively the needle holder) and paddle and push OK to open the user interface for Biopsy Unit referencing. After successful referencing the system should be restarted. OK 1. Click OK OK 2. The following window is displayed. Click Allow blocked content @ lixopsy Referencing Tael - Internet Explorer provided by'S a error has occurred while trying to access Biopsy referencing page. Internet Explorer restricted this webpage from running scripts or ActiveX controls. Allow blocked content x SIEMENS Healthineers Job Aid | How to Perform a New Biopsy Unit Referencing- Error 7 Effective Date:07/28/2021 | HOOD05162003207359 3. The following window is displayed. Click Yes e Biopsy Referencing Tool - Internet Explorer provided by Siemens Healthcare An error has occured while trying to access Biopsy referencing page . Windows Internet Explorer The webpage you are viewing is trying to close the window. Do you want to close this window ? NO 4. The Zigma Webserver will open. Follow the onscreen instructions for Referencing: ensure the biopsy unit is plugged in, system is in a state where movements are possible, no paddle is mounted, and no obstacles are on the table. Click Start Referencing. Zigma Page sent at 15:38:32 19.05.2021 Webserver System Version VF02E State E STATE BLOCKED Biopsy Service Please check following: - Biopsy is plugged - System is in state where movements (compression unit, rotation, ... ) are possible - No paddle mounted - No obstacle on the biopsy table Start Referencing 5. The needle holder will automatically move through several positions. The needle holder should stop moving in the upper left position. Click OK if the needle holder is located in the upper left position OR Click Continue if the needle holder is NOT located in the upper left position. Zigma Page sent at 15:39:29 19.05.2021 Webserver System Version VF02E State E_STATE_BLOCKED Back to Home Service Mode Ok: confirm the Reference Position. The needle guide should be in the top left inside position Continue: Reference Position is not ok Retry: Move to the reference position again RK Continue Retry SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 How to Perform a New Biopsy Unit Referencing- Error 7| Effective Date: 07/28/2021 | HOOD05162003207359 Healthineers Job Aid | How to Perform a New Biopsy Unit Referencing- Error 7 Effective Date:07/28/2021 | HOOD05162003207359 6. When referencing is successfully completed you will see “Referencing done!” Zigma Page sent at 15:41:08 19.05.2021 Webserver System Version VF02E State E_STATE_BLOCKED -Back to Home Referencing done! 7. Click the in the upper right corner of the window to close the application. 8. The biopsy unit is ready for operation. SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 How to Perform a New Biopsy Unit Referencing- Error 7| Effective Date: 07/28/2021 | HOOD05162003207359 Healthineers Job Aid | How to Perform a New Biopsy Unit Referencing- Error 7 Effective Date:07/28/2021 | HOOD05162003207359 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 How to Perform a New Biopsy Unit Referencing- Error 7| Effective Date: 07/28/2021 | HOOD05162003207359 Healthineers
- Error 7
- Biopsy
- Mammomat
- Revelation