Import Protocol Workflow

Import Protocol Workflow

Import Protocol Workflow  Note: When importing protocols the system will be blocked during this time. It is recommended to start a protocol import only when no scans or other applications are running. 1. Import protocol Exam Explorer - WUSER  Load CD/DVD in drive (or) insert flash drive Object Edit Tools Insert  Open Exam Explorer Open Ctrl+O  Select > User Tree Ctrl+N  Select > Object > Import Save Ctrl+S  Pop-up message appears > Select > OK Save AS ... Exam Explorer Import X Print ... Ctrl+P Export Import You are about to import protocols from a storage media. This action may take several minutes. Exit Please be aware, that the system will be blocked during this task. We recommend to start a protocol import only when no scans or other applications are running. Do you want to import now? OK Cancel 2. 3. Import Object box opens  Drives drop-down o Select > U Drive when using a CD/DVD (or) o Select > F or G Drive when using a flash drive Select >Import Import Object X File Name: Folder: T-SPINE ---. edx MRE tc.edx PEDIATRIC PROTOCOLS.edx research.edx SCMR CARDIAC AVANTO 32SQ.edx SCMR CARDIAC B17 VERIO_18-32 standard.edx Thalassemia.edx T-SPINE ---. edx TWIST_B15 PROTOCOLS.edx UPPER EXT-UTAH.edx File Type: Drives: Exam Database ( *. edx) g: NANO PRO c: Med_System f: MR *Pediatric: MR scanning has not been g: NANO PRO established as safe for imaging fetuses and Import Cancel i: Med IPT infants less than two years of age. The r. NECK t: Med_System responsible physician must evaluate the lu: Med System benefits of the MR examination compared to those of other imaging procedures SIEMENS www.usa.siemens.com/healthcare 703 Effective date 9/8/2016 Page 1 of 3 4. Select > File Name from box on left side  Note: File name will end in (.edx)  Note: Can only import one protocol at a time 5. Select > Import  Programs to be Imported box may open if conversion is needed o Protocols may need to be converted first o Select > Convert Programs to be imported x Import - Conversion needed WIR WUSER\T-SPINE -- \ROUTINE\ROUTINE Import Convert Cancel Help  Conversion completed > Select > Import  Note: Sometimes an Error Message may appear stating Import is not allowed. The protocol must be imported at the same level that it was exported. Below is an example of one message that may appear. Exam Explorer X x Import is not allowed. You are trying to import a region into an exam. OK www.usa.siemens.com/healthcare SIEMENS Page 2 of 3