Inspecting the Atellica® Sample Handler Robot Gripper O-Rings Video Version 1.25 and Below

This video provides instructions on how to perform the monthly maintenance task of inspecting the Atellica® Sample Handler robot gripper O-rings.

The Atellica sample handler robot gripper Orings should be inspected as part of monthly maintenance. To inspect the O rings 1st put the sample handler into either pause or stop mode, pausing the sample handler enables the system to complete all tests currently in process while stopping the sample handler stops the system immediately and discards all tests currently in process. To put the sample handler into pause or stop mode, first select the system button on the command bar. And then select the status tab. And the sample handler tab. On the sample handler status screen, select either the pause or stop icon. Once the sample handler is paused or stopped, open the sample handler front cover by first selecting the button to unlock the front cover. Wait for the lock symbol to change to an unlock symbol. Then open the front cover within 30 seconds of disabling the cover lock. If the cover is not opened within 30 seconds, it will automatically relock. Manually move the sample handler robot to the front of the sample handler. Do not hold the sample handler robot grippers to move the robot, or push the robot down. When moving the robot manually, only move the robot by holding the robot body and keep the robot grippers clear of all obstacles. If the robot is out of reach from the front of the sample handler, open the sample handler back cover to move the robot into position. To access the O rings manually, raise the gripper assembly and hold it in place. Visually inspect the four O rings for any signs of wear or damage. If any o-ring show signs of wear or damage, replace all four O rings. To replace the orings, first remove all of the racks from the sample drawers. Place paper towels below the robot grippers to catch any orings that may fall. Manually raise the gripper assembly and hold it in place to access the orings. Grasp and pull each of the four gripper orings off its gripper finger. Dispose of the old orings per laboratory protocol. Place a new o-ring on each of the four gripper fingers. Remove the paper towel from below. The sample handler robot gripper. Close the sample handler front cover if appropriate. Also close the sample handler back cover. Lock the sample handler front cover by selecting the button to lock the front cover on the sample handler status screen. If appropriate. Also lock the sample handler back cover. To complete the procedure, start the sample handler by selecting the resume icon on the sample handler system status screen. Mark, the task is completed in the maintenance schedule by first selecting the maintenance button on the command bar. And then select the Schedule tab. In module select sample handler from the dropdown menu. Select the activity. Inspecting or replacing SH. Robot gripper orings. And finally select Mark as completed to log the activity is completed in the maintenance log.

SIEMENS . SIEMENS. AtellicaTN Solution AtellicaTb' Solution AtellicaTN' Solution AtellicaTM' Solution AtellicaT'•' Solution AtellicaT•' Solution Version 1 , Version 1. Version Version 1.12.1.2846002 12.1.2846002 .12.1.2846002 SP00107 IH00104 Honlthinøør« Hønlthinøør« . Healthineers Setto InveN-ory Inver-ory Invaü»ry Invaü»ory InveNto»ry Invertory Invaütory Inveraory Inveradory Orders Worklist Workhst Maintenance Maintenarwe M üitenarx:e M amtenarx:e M üintenarx:e SettO Inventory InveMtory Status Operator Diagnostics Diagnostics Data Restore Software Updates Copy Configuration VMM p.p1-04 cC-P1-05 c-P1-05 pl-02 p 1-02 Module: All Frequency Daily Time. Next 24 hours Performed By All Module: All Frequency. Daily Time: Next 24 hours Performed By: All Module: SP00107 Frequency All Time: Next 24 hours Performed By: All Module: SP00107 Frequency All Time: Next 24 hours Performed By. All Module: SP00107 Frequency All Time: Next 24 hours Performed By. AJI System VMM -01 CH IA-03 All SP00107 Module Module Module Activity Activity Scheduled Sun-21 Mon-22 Tue-23 Wed-a Thu-25 Fri-26 Sun-21 Mon-22 Tue-23 Wed-24 Thu-25 Fri-26 Fri-26 Sun-21 Sun-2' Sat-27 Frequency All All Frequency Cleaning the SHC Fan Cleaning the SHC Fan Finer 1/1/2018 7:00 7:00 Mark as Standby Stopped Standby Standby Stopped Time Next 24 hours Next 24 hours Time CH Dairy Maintenance CH Daåy Maintenance SH Autocheck 1/24/2018 400 7:00 X Not Performed. Not Perform... Category InveNtory Tme Tirne Timne Time Tune Performed All All Performed Capabilities SPOOIII Cleaning the Sample Handler Drawer and the Racks Cleaning the Sample Handler Drav,er and the Racks As Needed 1/24/2018 1124/2018 7:00 Procedtre Details Procedure Detais Enter the activity name Enter the activity name. CM00113 Search Text Time CM00113 SP00107 Robot Robot 2 Robot 1 Inspecting the Washer Probes Inspecting or Replactng the SH Robot Gripper O-Rings Inspecting or Replacing the SH Robot Gripper O-Rings As Needed 1/24/2018 200 Cl TCS Scheduled by Spervtce Scheduled by Operator Database Backup 0:30 Scheduled Autornaticalty Cal-QC Storage Cal(2C Storage TCS SP00107 SH Autoc heck Scheduled by Service LAS SPOOIII 1M Autocheck SH Autoc heck SH Autocheck 1/25/2018 100 Connected to PCC A Failed Failed Drawer 1 Front Cover Back Cover VMM Atelhca Transport Autocheck Atellica Maghne Transport Autocheck Overdue Drawer 3 ox Inprocess Move to Historical Drawer 2 Drawer 1 Drawer 4 Drawer 3 Completed CM00113 Daily Cleaning IMT Daffy Cleaning Drawer 2 Lock Front Cover Cover X Not Performed Not Performed Front Cover Lock Front Cover Front 3300 Front Cover Back Cover With Comments Back Cover Front Cover 3:30 Autochec SP00107 1/25/2018 3:30 Add Activty Add Activiy Add Activay . Set Warning... Set Warning. Set Warning PEP LabM LabManaøer LabM anaøer LabManager LabM anager LabM3nager LabM3naøer Wednesday. January 24. 2018 2:45 PM Wednesday. January 24.2018 2:45 PM Wednesday, January 24.2018 2:45 PM Wednesday. January 24.2018 245 PM Wednesday, January 24.2018 2:452 PM Wednesday, January 24.2018 2: 125 PM Wednesday. January 24.2018 2:4513 PM Wednesday, January 24, 2018 2: 15 PM