MAMMOMAT Inspiration VB30: How to Transfer Images Manually
MAMMOMAT Inspiration VB30: How to Transfer Images Manually step by step
MAMMOMAT Inspiration VB60: How to Transfer Images Manually Women’s Health/Mammography 1. Find the desired patient in the Local Database. 2. To MANUALLY TRANSFER, go to the Series, and select the TOMO images that have the exten- sion STANDARD_SEGMENTATION (Ex: TOMO LMLO, STANDARD_SEGMENTATION) These are your slices Patient Transfer | Edit View Worklist Filter Sort Private Applications Options Help Import Ctrl+ Archive to DVDBURNER Archive to ... Send to Export to. . Eject from DVDBURNER Eject from. MAMMO SCREEN [13] TOMO L-MLO Standard_Segmen Finalize Medium end Eject from IDBURNER S [12] T_PRL-MLO. Screening EF Local Job Status. Ctrl+L D Network Job Status Ctrl+N [11] T PR raw L-MLO Screening comme x Go to the top dropdown menu bar and click on TRANSFER. Click on ARCHIVE TO, then click on PACS and click OK. 3. If you would like to transfer ALL 4 TOMO views, hold down the CONTROL key, and select each view in the series. (Ex: TOMO LMLO, Standard; TOMO RMLO, Standard; TOMO RCC, Standard; TOMO LCC,Standard) Go to the top dropdown menu bar and click on TRANSFER Click on ARCHIVE TO, then click on PACS, and click OK 4. To MANUALLY TRANSFER your FIRST VIEW to PACS, you will need to locate your FIRST view, which is your 0 degree image. SIEMENS siemens.com/healthineers Healthineers Job Aid | MAMMOMAT Inspiration VB60: How to Transfer Images Manually Effective Date: 12/18/2017 | HOOD05162002659842 5. Go to the series and locate the images that are labeled (Ex: T_PR_LMLO, Screening). Your First view ) degree image will be located here, and labeled as 0. Patent Transfer| Edit View worklist Elker sor rivate Applications Options Help X Archive to DVDRW Archive to .. Send to .. Export to ... [11] T_PR_raw L-MLO PRIME. Eject from DVDRW [10] TOMO R-MLO PRIME, Sta Eject from ... conVINI7 , Digital breast tor [9] T_PR R-MLO PRIME, Scree Finalize Medium end Eject from DVDRW [8] T_PR_raw R-MLO PRIME, Local Job Status ... Ctrl+L Network Job Status ... Ctrl+N [7] TOMO R.CC PRIME, Stand Import from Off-line ... [6] T_PR R-CC PRIME, Screen Export to Off-line ... Record Off-line Files [5] T_PR_raw R-CC PRIME.S( JAMES, NANCY, L [3] T_PRL-CC PRIME, Screen JARED, SHARAZO [2]T_PR_rawL-CC PRIME, Sc Ditarne JARNIGAN, BRENT ciption Digital breast tomosynthesis, bilateral Date of birth 6/23/1961 & Patient ID: 52325 Study dato 5/16/2017 Modality MG Go to the top dropdown menu bar and click on TRANSFER Click on ARCHIVE TO, then click on PACS, and click OK. 6. If you would like to transfer ALL 4 FIRST view images, hold down the CONTROL key, and se- lect each series labeled with this format. (Ex: T_PR_RCC, Screening) 7. To transfer your RAW data for your FIRST view to ICAD, you will need to locate the raw data in the series. It will be labeled as T_PR_raw LCC PRIME. Local Database HUMANN, CONNE Digital breast torL [2] T_PR_raw L-CC PRIME, Sc Scheduler HUNTER, DEBRA DVDRW HURLEY, KAY, C HURST, TERESA, ! JACKSON. CRYST JAMES, NANCY, L JARED, SHARAZO JARNIGAN, BRENT Patient name HUMANN: CONNIE. F. Date of birth 6/23/1961 Patient ID 5232 Study doscription Digital broast tomosynthesis: bilateral Study dato 5/16/2017 Modality The 0 degree image will be listed in the images. 8. Click on the 0 degree image, and then, from the top dropdown menu, click on TRANSFER, and then SEND TO. Select ICAD and click OK. Patient Transfer Edit View Worldlist Filter Sort Private Applicatio ? ? Import Ctrl+I X Archive to DVDRW Archive to. ? Send to ... Export to ... Eject from DVDRW Eject from .. Digital bre Finalize Medium and Eject from DVDRW com/ 12/ 29 Local Job Status .. Ctrl+L Network Job Status. Ctrl+N Siemens Healthineers Services Page 2 of 4 © Siemens Healthcare GmbH, 2017 Job Aid | MAMMOMAT Inspiration VB60: How to Transfer Images Manually Effective Date: 12/18/2017 | HOOD05162002659842 NOTE: DO NOT SEND ALL OF YOUR RAW DATA TO ICAD. ONLY SEND THE 0 DEGREE (FIRST IM- AGE). IF ALL IMAGES ARE SENT, IT WILL CRASH YOUR ICAD. 2 12 NO! Siemens Healthineers Services Page 3 of 4 © Siemens Healthcare GmbH, 2017 Job Aid | MAMMOMAT Inspiration VB60: How to Transfer Images Manually Effective Date: 12/18/2017 | HOOD05162002659842 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 4 of 4 © Siemens Healthcare GmbH, 2017
- MAMMOMAT Inspiration PRIME Edition
- Stereotactic Biopsy with MAMMOMAT Inspiration