MAMMOMAT Inspiration VB60D - How to Move an Auto QC Test to Completed
MAMMOMAT Inspiration VB60D - How to Move an Auto QC Test to Completed
MAMMOMAT Inspiration VB60D How to Move an Auto QC test to Completed 1. Once a QC test has been completed, locate SUBMIT REPORT under the performed Auto QC test. Equipment Checklist Test cases performed ID Test Case Tester Date Status INTRO Prerequisites LBIJL 2017.06.13 accepted repeat report remove QCUS01 Phantom LB 2017.06.13 passed repeat report remove Imaging Quality submit report 2. The completed test will now move to WHITE with the next scheduled date to perform the test. Inspiration Automated Quality Control System Annual Tests (USA) overdue 2017.01.02 Semi-Annual Tests (USA) overdue 2017.05.01 Mammography Equipment Evaluation (QC Manual) on-demand Spatial Resolution (QC Manual) on-demand >Mammography Equipment Evaluation (USA) on-demand Printer Check (USA) on-demand Weekly Tests (USA) scheduled 2017.06.20 > PHANTOM WEEKLY (USA) scheduled 2017.06.20 Quarterly Tests (USA) scheduled 2017.09.01 3. When the test is due it will move to the RED section in Auto QC Inspiration Automated Quality Control System Annual Tests (USA) overdue 2017.01.02 Semi-Annual Tests (USA) overdue 2017.05.01 Mammography Equipment Evaluation (QC Manual) on-demand Spatial Resolution (QC Manual) on-demand > Mammography Equipment Evaluation (USA) on-demand >Printer Check (USA) on-demand > Weekly Tests (USA) scheduled 2017.06.20 >PHANTOM WEEKLY (USA) scheduled 2017.06.20 Quarterly Tests (USA) scheduled 2017.09.01 SIEMENS Healthineers www.usa.siemens.com/healthcare HRT:HOOD05162002624076 Effective Date: 7/23/2017 Page 1 of 1