MAMMOMAT Revelation: How to Manually Transfer a Tomo Study with Insight 2D & Insight 3D
Step by step instructions on how to manually transfer a tomo study with Insight 2D and Insight 3D on the MAMMOMAT Revelation
How to Manually Transfer a Tomo Study with Insight 2D & Insight 3D MAMMOMAT Revelation To manually transfer a tomo study you must follow each of the steps carefully. Failure to do so may result in images transferring to the incorrect destiantions. (Ex. If slices are sent to CAD it will CRASH CAD.) To MANUALLY TRANSFER TOMO SLICES, FIRST VIEW, INSIGHT 2D, and INSIGHT 3D to PACS: 1. Find the desired patient in the Local Database Local Database Test, Orange MAM 3D Screening Bi /com/ /a/ / / / 2. Select the TOMO views from the “SERIES” that have the extension EMPIRE -Ex: TOMO LMLO, EMPIRE. These are your SLICES. MAM 3D Screening Bi [4] TOMO R-CC PRIME, EMPIRE, Screening 1/com/ /a/ / / / /com/ / / / / / To archive ALL TOMO views at the same time, hold down the CONTROL key, and select each view in the “series” (Ex: TOMO LMLO, EMPIRE, TOMO RMLO, EMPIRE, etc) 3. Go to the top dropdown menu bar and click on TRANSFER, click on “ARCHIVE TO”, then click on PACS, and click ARCHIVE Archive To Destination Compression |Quali ... |Label on medium ImagoGrid Locclocc 100 PACS None 0 Transfer Edit View Import Archive to DVDRW Archive to .. Archive Cancel Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 SIEMENS Master Template: EFF Date 22 Aug 17 | HOOD/HILS HOOD05162002647849 Healthineers Jobaid: Effective Date: XX/XX/XXXX | HOOD/HILS Job Aid | How to Manually Transfre a Tomo Study with Insight 2D and 3D Effective Date: 05/03/2019 | HOOD05162003009739 4. Select the FIRST VIEW images from the “SERIES” labeled Ex: T_PR LMLO, MAM 3D Screening B [10] T_PR L-MLO PRIME, Screening Screening. w/com/ /a/ / / / To send ALL FIRST views at the same time, hold down the CONTROL key, and select each view in the “series” (Ex: T_PR LMLO, Screening, T_PR RMLO, Screening, etc) 5. Go to the top dropdown menu bar and click on TRANSFER, click on “ARCHIVE TO”, then click on PACS, and click ARCHIVE Archive To Destination Compression |Quali ... Label on medium ImagoGrid acclacc 100 PACS None 0 Transfer | Edit View Import Archive to DVDRW Archive to ... Archive Cancel 6. Select the INSIGHT 2D images from the “SERIES” labeled Ex: INSIGHT 2D, LMLO, MAM 3D Screening Bi [5] INSIGHT 2D R-CC PRIME, Screening Screening "/com/ /a/ / / / 'com/ // / / / To send ALL INSIGHT 2D views at the same time, hold down the CONTROL key, and select each view in the “series” (Ex: INSIGHT 2D LMLO, Screening, INSIGHT 2D RMLO, Screening, etc) 7. Go to the top dropdown menu bar and click on TRANSFER, click on “ARCHIVE TO”, then click on PACS, and click ARCHIVE Archive To Destination Compression |Quali ... Label on medium ImagoGrid Lacclacc 100 PACS None o Transfer Edit View Import Archive to DVDRW Archive to ... Archive Cancel Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 2 of 5 Job Aid | How to Manually Transfre a Tomo Study with Insight 2D and 3D Effective Date: 05/03/2019 | HOOD05162003009739 8. Select the INSIGHT 3D images from the “SERIES” labeled Ex: INSIGHT 3D, LMLO, Bilateral Tomo + 2D P [11] INSIGHT 3D Left-CC PRIME, Diagr Screening. ¿com/ /// / / DONE /com///// / COM/LOC To send ALL INSIGHT 3D views at the same time, hold down the CONTROL key, and select each view in the “series” (Ex: INSIGHT 3D LMLO, Screening, INSIGHT 3D RMLO, Screening, etc) 9. Go to the top dropdown menu bar and click on TRANSFER, click on “ARCHIVE TO”, then click on PACS, and click ARCHIVE Archive To Destination Compression | Quali ... Label on medium ImagoGrid acclacc 100 PACS None 0 Transfer | Edit View Import Archive to DVDRW Archive to ... Archive Cancel To send ALL Tomo, First View, INSIGHT 2D, and INSIGHT 3D views at the same time, hold down the CONTROL key, and select each view in the “series” (Ex: TOMO LMLO, EMPIRE, T_PR LMLO, Screening, INSIGHT 2D LMLO, Screening, INSIGHT 3D LMLO, Screening, etc) [22] INSIGHT 2D Left-CC, Diagnosis [21] TOMO Left-CC, Empire, Diagnosis ][20] T_PR Left-CC, Diagnosis [19] T_PR_raw Left-CC, Diagnosis foom/ //// / COM/LOC [18] TOMO Left-MLO PRIME, EMPIRE_C, Diagnosis [17] T_PR Left-MLO PRIME, Diagnosis ][16] T_PR_raw Left-MLO PRIME, Diagnosis joom/ //// / COM/LOC [][15] INSIGHT 2D R-MLO PRIME, Diagnosis | [14] TOMO R-MLO PRIME, Empire, Diagnosis LAmit!!: { COMILOC ][13] T_PR R-MLO PRIME, Diagnosis [12] T_PR_raw R-MLO PRIME, Diagnosis 7com/ 1111 1 COM/LOC [11] INSIGHT 3D Left-CC PRIME, Diagnosis (][10] INSIGHT 2D Left-CC PRIME, Diagnosis Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 3 of 5 Job Aid | How to Manually Transfre a Tomo Study with Insight 2D and 3D Effective Date: 05/03/2019 | HOOD05162003009739 10. Go to the top dropdown menu bar and click on TRANSFER, click on “ARCHIVE TO”, then click on PACS, and click ARCHIVE Archive To Destination Compression Quali Label on medium IlmagoGrid LLacclacc PACS None 0 Transfer Edit View Import Archive to DVDRW Archive to ... Archive Cancel To MANUALLY TRANSFER “RAW” FIRST VIEW to CAD: This step CANNOT be combined. Each view must be done individually. 11. Select the RAW FIRST VIEW images the “SERIES” labeled Ex: T_PR_raw LMLO, Screening. MAM 3D Screening Bi [6] T_PR_raw L-MLO PRIME, Screening I/com/ /a/ / / / 12. Click on the 0 degree image at the end of the 25 projections in the thumbnails. Patient Tame Test, Orange Study description MAM 3D Screening Bilateral Date Of DITLIT 270/2019 Study date 2/6/2019 . . 13 12 11 10 9 8 7 6 5 4 1 13. Go to the top dropdown menu bar and click on TRANSFER, click on “SEND TO”, then click on iCAD, and click SEND Send To pression Quality fvh icad Nole Transfer Edit View Worklist Import Archive to DVDRW Archive to ... Send to ... Send Cancel T Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 4 of 5 Job Aid | How to Manually Transfre a Tomo Study with Insight 2D and 3D Effective Date: 05/03/2019 | HOOD05162003009739 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (herein- after collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2018 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 5 of 5
- MAMMOMAT Revelation: How to Manually Transfer a Tomo Study with Insight 2D & Insight 3D