MAMMOMAT Revelation How do I Know When the Biopsy Calibration is Needed
MAMMOMAT Revelation How do I Know When the Biopsy Calibration is Needed
MAMMOMAT Revelation How Do You Know if Biopsy Calibration is Overdue? 1. Turn on the MAMMOMAT Revelation 2. A pop-up message box will display WH AWS Biopsy calibration was not performed for the mounted type of biopsy table. Targeting accuracy cannot be guaranteed. Image acquisition is not possible. Please mount a different kind of biopsy table or perform a biopsy calibration. OK 3. Perform Biopsy Calibration SIEMENS Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Healthineers Job Aid | How Do I Know When Biopsy Calibration is Needed? Effective Date: 06/25/2020 | HOOD/HILS05162003104973 Job Aid | How Do I Know When Biopsy Calibration is Needed? Effective Date: 06/25/2020 | HOOD/HILS05162003104973 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Job Aid | How Do I Know When Biopsy Calibration is Needed? Effective Date: 06/25/2020 | HOOD/HILS05162003104973 Healthineers Page 2 of 2
- MAMMOMAT Revelation