MR Cardiac Flow Numaris X

BMR MR Cardiac Flow Numaris X The MR Flow Quantification step is used to determine flow and velocity parameters. 1. From the Patient Browser, load the appropriate data set into the MR Cardiac Analysis Workflow A. Rght-click on the patient examination and select Assign Workflow > MR Cardiac Analysis B. Double Left Click on Patient to open 2. The Cardiac Reading Step enables you to view functional, dynamic, tissue characterization and flow data using the common tools 3. Select the MR Flow Quantification Step MR Cardiac Analysis + MR Cardiac Reading MR Cardiac Function MR Flow Quantification 4. Perform Vessel Segmentation: Select an image with flow Vessel Segmentation O Backward Forward A. Select > ROI1 to activate drawing an ROI Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 SIEMENS Job aid: Effective Date |#HOOD05162002981764 Healthineers Job Aid | MR Cardiac Flow Numaris X Effective Date01/02/2019 | HOODHOOD05162002981764 B. Select > one-click vessel segmentation icon and click in the vessel or use one of the drawing tools to draw an ROI C. Use the vessel segmentation tools to refine the segmentation, if required D. Scroll through the phases to check the segmentation E. Select > Copy Forward or Copy Backward icon Vessel Segmentation O O. O Per Nuage Adjuil F. If you want to perform other flow evaluations in a slice position, click the ROI2-ROI4 icons to enable the drawing of other contours G. In the lower left segment, the results of the vessel segmentation are displayed as a time-flow graph and a result ticker: oThe time-flow graph displays the flow quantification results over time for the selected flow parameter. These are the following flow parameters to be displayed:  Flow: Product of mean velocity and surface area  Peak-velocity: Peak velocity in the ROI  Vel: Mean velocity in the ROI  Net Flow: Difference between forward and reverse flow  Area: Surface of the segmented vessel Flow Vel Net Flow Aroc ROM RO13 ROH cm2 1.0 3.7 31 2.8 2.5 2.2 1.6 1.3 1.0 10 20 30 40 50 60 70 80 90 100 107 214 321 128 535 641 748 855 1069 Start End Peak-Vel Forw-Volume Rey-Volume Avg-Flow/Min Ron -127.734 0.066 76.183 -4.11 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 2 of 6 Job Aid | MR Cardiac Flow Numaris X Effective Date01/02/2019 | HOODHOOD05162002981764 oSelect a flow parameter and select one or more ROI oClick and drag the start and end sliders to the desired position to adapt the time interval 5. In the Evaluation Options, step, you can optimize your evaluation results A. VENC Anti-Aliasing: If the measure maximum flow velocity is lower than the actual flow velocity, there will be signal void areas in the vessel due to aliasing MR Flow Quantification Vessel Segmentation Evaluation Options Venc Anti Alias. Background Phase Adjustment oSelect > VENC Anti-Aliasing icon, images are update. The aliased pixel values are corrected and the results are update oAdjust the background phase > Move the threshold slider to adapt the sta- tionary tissue oIf necessary, select > Auto button to restore the default stationary tissue mask oSelect > Exclude Regions icon and draw a contour around the regions that should not be included into the stationary tissue mask oSelect > Delete Exclusion Regions icon if you want to delete and redraw the exclusion region o Select > Apply to correct the pixels for the phase contrast images 6. For MR Flow Quantification, the results of the current quantitative analysis are available in Findings details A. Select the Findings details icon > a Findings details dialog box will open Findings Navigator Name Value Source B. Can define the heart rate of the patient. By default, the average heart rate is dis- played. An adaptation of the heart rate influences the average flow per minute. C. If you have performed the background phase adjustment, the check box in the Evaluation Option area is selected D Select > Invert Polarity icon to invert the velocity and flow values. E Select > Create Finding to create a finding for the selected ROI (repeat step for each ROI created) Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 3 of 6 Job Aid | MR Cardiac Flow Numaris X Effective Date01/02/2019 | HOODHOOD05162002981764 Findings details Patient information Evakantion Options Und: 154.10 LEL Frpale Finding Flow Average Velocity Arg Flow Per Ponk Velocity 314.32 HS 147 467 mm 115 5 ms Ponk Average Velocity Tiene lo Peak Average Veloceily Reverse Volume Pressure Cead ant [ avmach) Nel Formara Volume -70 117.5 Body Surtaxe Area 1.224 cm2 2 271 Mining Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 4 of 6 Job Aid | MR Cardiac Flow Numaris X Effective Date01/02/2019 | HOODHOOD05162002981764 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instruc- tions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2018 ………………………………………………………… Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 5 of 6 Job Aid | MR Cardiac Flow Numaris X Effective Date01/02/2019 | HOODHOOD05162002981764 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 6 of 6