MR Neuro 3D - Tractography - Numaris X JobAid
MR Neuro 3D - Tractography - Numaris X JobAid
MR Neuro 3D – Tractography Tractography Step – can be used to graphically display neural tracts. The Tractography step offers tools and to interactively create and evaluate diffusion tracts, or to statically create and visualize diffusion tracts. Tools for fused visualization fo the diffusion maps are also avilable, such as b0, ADC, FA, TraceW (trace-weighted), RD (radial diffusivity), and AD (Axial diffusivity) maps. Note: A Tensor dataset must be available to perform Diffusion Tractography 1. From the Patient Worklist, load the appropriate data set into the MR Neuro 3D Workflow a. Examination correctly mapped: double click to load b. Manual mapping required: right click the examination and Open With > MR Neuro 3D. O Open Ctrl+O @ Open with Favorites View as Read-Only with Suggested Batch Open MR Breast Assign Workflow MR BreVis Group Studies MR Cardiac Analysis Batch Assign Workflow MR Neuro 3D Cancel Workflow MR Neurology Search Patient with DICOM Q/R MR Oncology Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 SIEMENS Master Template: EFF Date 22 Aug 17 | HOOD/HILS HOOD05162002647849 Healthineers Jobaid: Effective Date: 02/07/2020 | HOOD/HILS HOOD05162003063274 Job Aid | MR Neuro 3D – Tractography syngo.via Effective Date: 02/07/2020 | HOOD/HILS HOOD05162003063274 2. Diffusion Maps – fused visualization maps H a. B0 fMRI 03, Course Siemens Healthcare MR 1232334 b. ADC *7/18/1982, M. 26Y Verio MR B15V CURRENT HFS c. TraceW – Trace Weighted 7/18/2008 t1 mprins sag_p2 iso d. FA – Fractional Anisotropy latest -- > Current.MR FUSION STRUCTURE STUDY ID 1 2:03:46 PM e. AD – Axial Diffusivity maps 13 IMA ~56 f. RD – Radial Diffusivity maps 5cm T1 900.0 TR 1900.0 Structure TE 2.5 MPR FUSION : p2 M/ND/NORM Function A1/IR 111211 ns / 9 Diffusion Tracts bo agent Diffusion Map ADC fMRI 03, Course 1232334 TraceW Siemens Healthcare MR *7/18/1982. M. 20% Verio FA MR B15V CURRENT HFS 7/18/2008 3. MR Neuro 3D Reading Step t1_mpr_ns sab p2 AD latest -- > Current.MR FUSION a. Image Selection Step RD 2.09.40 PM i. Structure – select from dropdown list 1. Dataset displayed in background of an MPR/VRT segment ii. Field Map – select from dropdown list (if available) 1. Select “No Field Map” if you don’t want this displayed 2. Note – Field maps are only available if no image alignent was used iii. Diffusion – select Tensor data from dropdown list MR Neuro 3D + MR Neuro 3D Reading Alignment Image Selection Structure t1_mpr_ns_sag_p2_iso.latest Field Map No Field Map Diffusion ep2d_diff_mddw_20_p2_TENSOR.latest Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 2 of 7 Job Aid | MR Neuro 3D – Tractography syngo.via Effective Date: 02/07/2020 | HOOD/HILS HOOD05162003063274 4. MR Neuro 3D Reading > Tractography Layout 1) MPR Segments 2) VRT Segment 3) Creating and Saving Tract bundles Icons 4) VOI range icons for selecting Boolean filter types 5) Tractography icons: Tractography VOI, Tractography Shrink to Activation; or Tractography Plane 6) Activating VOI Creation and Editing 7) Diffusion tract bundle settings list 8) List of created diffusion tract bundles MR Neuro 3D 9 + Tischagresky CURRENT (8) IMRI_03, Course Siemens Healthcare MR 1232334 "7/18/1982, M, 26Y (7)- MR BISV CURRENT 7/18/2008 (6) 11_mpt_ns_sag_02_iso latest -> Current.MR_FUSION_STRUCTURE STUDY ID 1 (5)- 2:03:46 PM (4) VOLT VOL_2 ( 3) MPR Images ALH Findings Navigator EX 10cm LP 11 900.0 Tools TR 1900.0 TE 2.5 VRT : p2 M/ND/NORM SPL109 ALAR "13di_ns / 9 Sage Cor(-30) Tra(8) VRT 3.45 AM 5. Tractography Step – Create Diffusion Tracts Interactively – create and evaluate diffusion tract bundles and view diffusion maps a. Tractography step is active MR Neuro 3D b. Diffusion Tensor data must be available MR Neuro 3D Reading c. Case Navigator > eye icon displayed if All Diffusion Alignment tracts displayed mage Selection i. Note: if no diffusion tracts have been defined Tractography “All Diffusion Tracts ” is selected by default AL Dimenon Tracts Sellinga Defaul. VOI 1 VOI 2 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 3 of 7 Job Aid | MR Neuro 3D – Tractography syngo.via Effective Date: 02/07/2020 | HOOD/HILS HOOD05162003063274 d. Select Static Tracking Icon – activated interactive track creation Static Tracting i. Select VOI or Freehand ROI icon > draw around region of interest ii. Shrink to VOI Icon – shrinks activation around region of interest iii. Plane Icon: draw a plane on the region of interest 1. Adjust position, size, and location of plane 2. Press Esc key or Enter key to finish edits 3. Note - Drawn VOI, Freehand ROI, or Plane defines seed point; tracts within drawn VOI are displayed. Non-selected tracts are hidden; displays are updated in all segments. Move the VOI or plane to adjust tract displays. iv. Draw additional VOIs, Freehand ROI’s or Planes to define further seed points v. Select 2 VOIs from the list (e.g., VOI 1 and VOI 3) 1. Tracts between 2 selected VOIs are displayed and Boolean type filtering selected is applied to these 2 VOIs. 6. Boolean Filters – select and use filters to display specific tracts a. A or B: Boolean filter type i. Diffusion tracts that run through VOI A or VOI B are displayed b. A and B: Boolean filter type i. Diffusion tracts that run through VOI A and VOI B are displayed c. A not B: Boolean filter type i. Only diffusion tracts that run through VOI A are displayed. Diffusion tracts that run through VOI B are not displayed d. B not A: Boolean filter type i. Only diffusion tracts that run through VOI B are displayed. Diffusion tracts that run through VOI A are not displayed e. Group Seed Point Objects icon (Lock): group two seed points linked with one of the previous Boolean filter types A or B A and B A not B B not A Group Seed Point Objectives Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 4 of 7 Job Aid | MR Neuro 3D – Tractography syngo.via Effective Date: 02/07/2020 | HOOD/HILS HOOD05162003063274 7. Save Diffusion Tractography > select one of the following a. Calculate & Save icon: create and save new tracts with selected algorithm and visual settings. Selected tract bundle is overwritten b. Calculate and SaveAs icon: create new tracts with selected algorithm and visual setting and save them as a new tract bundle. Use to create a new bundle based on an existing one. c. Save As It Is icon: display tracts in image area as is. New diffusion tract bundle is displayed with name of the algorithm parameter preset selected from settings list. Calculate and Save Calculate and Save Calculate and SaveAs Save As It Is 8. Edit and Delete Tracks a. Edit Track MR Neuro 3D Q + fMRI 03, Course i. Tractography Step 1232334 Tractography *7/18/1982, M. 26Y ii. Right mouse > select Diffusion CURRENT All Diffusion Tracts 7/18/2008 Tract Bundle created > select t1 mpr_ns_sag_p2 Default_1 Edit Rename Rename 1. Enter New Name for Delete Diffusion Tract Settings Properties Tracting Default Statistics Create Neuro 3D Finding b. Delete Tract Bundles i. Tractography Step ii. Right mouse > select Tract created > select Delete iii. Select Yes – confirms deletion MR Neuro 3D G + fMRI 03, Course 1232334 Tractography *7/18/1982, M, 26Y CURRENT Confirm AF X All Diffusion Tracts 7/18/2008 t1_mpr_ns_sag_p2 Default_1 Edit ? Do you want to remove the selected diffusion Rename tract bundle? Delete Settings Properties Static Tracting Default Statistics Yes No Create Neuro 3D Finding Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 5 of 7 Job Aid | MR Neuro 3D – Tractography syngo.via Effective Date: 02/07/2020 | HOOD/HILS HOOD05162003063274 c. Delete Tract Fibers i. Image Segment > Right mouse > select Tract Deletion Mode ii. Click on Fiber to Delete (e.g., fiber is white) iii. Note: if Tract Deletion Mode isn’t selected and you pick a fiber, the MPR segments displays the selected 3D point on the surface of the fiber bundle IMRI_03, Course Siemens Healthcare MR 1232334 "7/18/1982, M. 26Y AR B15V CURRENT 7/18/2008 11_mpr_ns_sag_p2_iso latest -> Current.MR_FUSION_STRUCTURE STUDY ID 1 2:03:46 PM Paste Measurem Tract Deletion Mode LIP TI 900.0 TR 1900.0 TE 25 VRT : p2 MIND/NORM SP L10 9 AMAR "t3d1_ns / 9 Sag>Cor(-39). Tra.es 9. Save Results Save and Send: Results are stored on the system and sent to your archive II Save and Pause: Results are stored on the system Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 6 of 7 Job Aid | MR Neuro 3D – Tractography syngo.via Effective Date: 02/07/2020 | HOOD/HILS HOOD05162003063274 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2018 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018 Page 7 of 7
- MR Neuro 3D
- Tractography
- Numaris X