PEPconnect

Multitom RAX: Performing a Real 3D

Performing Real 3D Job Aid Multitom Rax 1. Open the Patient in the Examination tab on the FLC. SERVICE uGym" Examination Flunro Rad Service, Service Service Fluoro Physics Serles 3D 15 60 C 0.0 MAS 0.0 me min 000.0 Dose 0.70 0.0 mr. Ou 0,2 mm Cu 0.1 30 0/3 Focus mymin %=70+ 0.0 0.0 Select orgen program 2. Select the desired 3D OGP View for the exam If necessary, make the suggested table adjustment Select the image laterality Verify the correct grid is inserted atient Position / View Position Please define the patient position View Position Please define the image laterality Head First Feet First Err 5548 / 240: Insert grid with 115 cm H Left Right None OK Cancel Please define the image laterality OK Left Right None OK Cancel Note: (a.) For Standing 3D exams, place the upright 3D stand into the pre-set position before moving the system into place. (b.) For Table-side upper extremities, move the system into place before attaching the arm board to the table. SIEMENS Siemens.com/healthineers Healthineers Performing Real 3D Job Aid | Multitom Rax Effective Date: 01/11/2018 | HOOD05162002769296 3. Keep the OGP button pressed on the hand-held remote control to move to the Start Position of the 3D Run. i Keep OGP button pressed to move to the start position of the 3D run 4. Make any necessary anatomy adjustments and use the hand-held remote control to set the isocenter of the anatomy in the first position. 5. Press the lower right violet key on the hand-held remote control to move to the second centering position. i Press lower right violet key at the remote control 3D-test-run active. Please to start the 3D-test-run move forward to end position 6. Make any necessary anatomy adjustments to set the isocenter of the anatomy in the second position. 7. Continue to press the right lower violet key on the hand-held remote control the system will stop once the end position is reached. i 3D-test-run active. Please move forward to end position Siemens Healthineers Services Page 2 of 5 Performing Real 3D Job Aid | Multitom Rax Effective Date: 01/11/2018 | HOOD05162002769296 8. Return to the start position by pressing the lower left violet button on the hand-held remote control. i Move back to start pos. by pressing lower left violet button at remote control 9. The system will stop at the first required position Engage fluoro in this position (remember that the system will stop the fluoro automatically). i Invoke fluoro to acquire parameters and wait till fluoro stops automatically 10. Press the lower left violet button on the hand-held remote control. i Move back to start pos. by pressing lower left violet button at remote control 11. The system will stop in the second required position. Engage fluoro in this position (remember that the system will stop the fluoro automatically). 12. Press the lower left violet button on the hand-held remote control. i Invoke fluoro to acquire parameters and wait till fluoro stops automatically Siemens Healthineers Services Page 3 of 5 Performing Real 3D Job Aid | Multitom Rax Effective Date: 01/11/2018 | HOOD05162002769296 i Move back to start pos. by i pressing lower left violet button at remote control Position reached. Release button at remote control 13. Engage the Acquisition to perform the imaging. Note: if you prematurely release the acquisition you will abort the spin. i i Press Acquisition switch to 3D run finished. Want to start the 3D run repeat: Then move system first back to start position 14. 3D images are automatically transferred to the 3D workstation for reconstruction. Siemens Healthineers Services Page 4 of 5 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018