MULTIX Impact: Defining the Order of OGP's within an Exam

This job aid will cover the steps to follow when setting the order of OGP's within an exam.

MULTIX Impact: Defining the Order of OGP’s within an Exam XPRF By default, Organ Programs (OGP’s) are listed in alphabetical order. The order of OGP‘s within an exam is customizable by using the following process. Note: “View Position” is the position of the patient during an exam. In this example, the system displays the views for a 2 View Chest in alphabetical order. The Lateral View is followed by the PA View. OGP List Patient Chest LAT Chest PA Mode Examination Patient Size AEC Postproc S 1) On the “Service” tab select the “Exam Properties” icon under “User Settings”. User Setting Site Info. General Processing Exam Setting Properties Properties Image Register DICOM Disk Display Config. Properties Manager User Manager Service PMS Remote Remote DICOM Assistance Access Node FD Manager Sensitivity SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 07/06/2020 | HOOD05162003107067 Healthineers Job Aid | Defining the order of OGP Effective Date: 07/06/2020 | HOOD05162003107067 2) In the “View Position” column on the right, move the “View Position(s)” in the desired dis- play order using the up and down arrows. Note: The order of the “View Positions” listed will apply to ALL exams. Meaning, the PA view is displayed first for a chest, hand, tib/fib, etc. followed by the AP, then the obliques, etc. Examination Properties X General Show Predefined Demo Image Auto Accept Image Heat Unit Test Procedure Order Body Part Protocol View Position Skull PA Neck AP Shoulder LLO Spine RLO Thorax LL Abdomer RL Pelvis LLD UpperLimb RLD Hand Not defined LowerLimb Foot OK Apply Cancel 3) Select “Apply” and then “OK” to save the update. 4) To assign each OGP a “View Position”, select the “OGP Editor” icon from the “Tools” section on the “Service” tab. Tool OGP Editor Clean FD FD Archiv ... Calibration Reconnect RIS Mapping Statistics Audit Trail Test Image SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Page 2 of 5 Job Aid | Defining the order of OGP Effective Date: 07/06/2020 | HOOD05162003107067 5. Select the “Organ Program” tab in the “OGP Editor”. OGP Editor X Procedure Organ program Configuration Search Body part All OGP list Renal AP Attribute X-Ray Parameter Image processing parameter ID Renal AP Acq type Default ID : Abdomen Table Name : Abdomen AP . Name Renal AP Wall Flavors Cassette Description Renal AP Default Free Name Standard . Acq mode DR RAD Exp mode : 1pt Smooth Body part Abdomen High contrast . Focus Finely detailed Anatomic ... Abdomen ID : T_Renal AP Customized Image Lat ... Not defined Name : T_Renal AP Image flip ... SID : 115 Parameters Name Value Minimum Maxdmum Image rot ... . 90 . 270 Grid type : 115 Contrast 2.3 0.5 5 Image label . Filter type : 0.0mm Detail 1.1 8.1 Latitude View posit ... Not defined Collimator center : Center Noise 0.1 WLShift Target EXI 250 Colimator length : 17IN(43CM) wwshift Viewicon Umage\Position\Human\Siemensts .. Collimator width : 17IN(43CM) Edit Remove Edit Print New Save Delete Cancel 6. For example, select “Thorax” from the “Body part” list using the drop down arrow to filter the OGP’s. Select “Chest PA” from “OGP list”. OGP Editor Procedure Organ program Configuration Search Body part Thorax OGP list Chest PA Attribute X-Ray Para Chest PA Chest OBL Right ID Chest PA Wall Chest OBL Left * Name Chest PA Cassette Chest AP Supine Free Chest PA Chest AP Upright Description Exp Chest Apical * Acq mode DR RAD Chest Decub LAT Body part Chest Dorsal Decub horax Chest Ventral Decub Anatomic ... Chest Ribs AP Bilateral Upper Name : W_Chest PA Image lat ... Not defined SID : 180 Image flip ... ..... SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Page 3 of 5 Job Aid | Defining the order of OGP Effective Date: 07/06/2020 | HOOD05162003107067 7. Under the “Attributes” column locate the “View position”. Use the drop down arrow to update the “View Position” by changing from undefined to PA. Select “Save”. OGP Editor X Procedure Organ program Configuration Search Body part Thorax OGP list Chest PA Attribute X-Ray Para Chest PA Image processing parameter Chest OBL Right . ID Chest PA A ID : Chest AP Chest OBL Left Wall Name : Chest PA . Name Chest PA Cassette Chest AP Supine Flavors Free * Description Chest PA Chest AP Upright Name Default Exp Chest Apical Standard Acq mode DR RAD Chest Decub LAT S Smooth Body part Thorax Chest Dorsal Decub High contrast Chest Ventral Decub Finely detailed Anatomic ... Chest Ribs AP Bilateral Upper Customized Name : W_Chest PA Image lat ... Not defined SID : 180 Image flip ... Parameters Grid type : Universal ..... Name Value Minimum Maximum Image rot ... 90 . 180 . 270 Filter type : 0.0mm Contrast 1.3 0.5 Image label R Detail 1.1 8.1 Collimator center : Center Latitude 10 View posit ... PA Noise 0.1 Collimator length : 17IN(43CM) WLShift Target EXI 250 Collimator width : 17IN(43CM) WWShift -8 O View icon Umage\Position\Human\SiemensS .. Edit Remove Edit Print New Save Delete Cancel 8. Repeat the same process for the Chest LAT OGP assigning a “View Position” of LAT. Note: The same process will need to be repeated for any exam where the OGP’s are not in the desired order. SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Page 4 of 5 Job Aid | Defining the order of OGP Effective Date: 07/06/2020 | HOOD05162003107067 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter co llectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written auth ority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Page 5 of 5