MULTIX Impact C & VA20: FD Calibration
This video will guide users through the proper steps to calibrate the Flat Detector (FD)
This video will demonstrate how to perform FD calibration to perform FD calibration. Ensure that you're logged in with a user account which has the FD calibration permission. Then access the service mode of the image station. Move the tube, stand away from the table top and position the tube pointing towards the floor. Placed the detector on the floor in the landscape position. Set the tube to maximum ID which should be at least 150 centimeters, but 180 centimeters is recommended. Then use the retractable ruler to measure the set side. Be sure to make a note of the side as it will be needed for calibration. Center the detector in the beam field and collimate the beam field to fully cover the active detector area. Insert the 0.6 millimeter copper filter. Click FD calibration. Please wait a moment while the system verifies the detector parameters and warms up. Enter the measured side in the current value box and click OK. A confirmation dialog appears. Ensure that the 0.6 millimeter copper filter has been inserted and click OK. A confirmation dialog appears. Turn on the calibration switch and click OK to continue. The calibration switch can be found on the backside of the generator on off console. Ensure that nobody is in the examination room and click begin to start automatic. X Ray release. The calibration is then started. Warning, do not turn off the calibration switch when the calibration is in progress. Wait until the calibration is finished and then another confirmation dialog will appear when the calibration is complete. Click OK to save the calibration results. Wait until the calibration results have been saved successfully. A confirmation dialog will appear as an indication that the results have been saved. Then click OK and reboot the entire system to activate the new files. Lastly, it is very important to turn off the calibration switch for normal operation of the system or else radiation is not possible. If for some reason the calibration does not succeed, the previous calibration remains valid.
SIEMENS Healthineers Archiving Ar&iving A'åhiving Detector Calibration Search Cali Dose nGy nGy Dose nGy Dose n Gy Dose nGy Dose Dose n Gy RIS FD Calibration Calibration ON MS Mappülg Mappülg Mapping Statistks Result List No Reference Image No Reference Image No Reference Imnage No Reference Image No Reference Image No Reference Image No Reference Image No Reference Image NO Reference Image NO Reference Image No Reference Image NO Reference Image NO Reference Image NO Reference Imnage NO Reference Image NO Reference Image No Reference Image No Reference Image No Reference Iin-lage No Reference Image No Reference Image No Reference Image No Reference Image No Reference Im-lage No Reference Image sps Description Age Setting Dose 37500 n Gy 37500 n Gy Dose 37500 n Gy Dose nGy Dose n Gy 31500 nGy Dose 37500 n Gy 37500 n Gy Dose 37500 n Gy Dose nGy Dose n Gy Dose n Gy SIEMENS that are in the beam path wait while the system that no additional are in the beam path wait while the system Confirmation Ionformation Information Ma Cali Ca la pa Density Sensitivity Current value Rea Current Confirmation Ionformation No Reference Image No Reference Image NO Reference Image NO Reference Image NO Reference Imnage No Reference Image No Reference Image No Reference NO Reference Image NO Reference Image No Reference Image No Reference Image No Reference Image NO Reference Image NO Reference Image NO Reference Imn-lage No Reference Image No Reference Image No Reference linage No Reference Image No Referenc NO Reference No Reference No Reference Image No Reference Image No Reference Image and tum it to in field to in and to in held and to the area. tube to the tube to the area. the to the the to the area. set to SID (at 150 Acquiring Healthineers ' the grid NO YES 10min nGy Dose n Gy Dose nGy nGy n Gy nGy Dose isooo nGy Dose 37500 n Gy Dose 'SOOO nGy Dose nGy Ca Cu bigw Cali Sensitivity Tube NO Reference Image NO Reference Image No Reference NO Reference Cali Ma Display all Total count Total count: 10% HU 13% HU Tube stand is in free position. Tube stand is in free position. 12141 Tube stand is in free position. 12/41 Calibration ON Calibration ON 12/21 Calibration switch; ON 12/21 Calibration switch ON 12121 Calibration ON 12121 Calibration switch; ON No IS not ready FD Calibration patient Dose No IS not No
- multix Impact
- multix impact C
- va20