MULTIX Impact C & VA20: Modifying Default Exposure Parameters
This video will cover the steps to modify default Exposure Parameters on the system.
This video will illustrate how to modify default exposure parameters. In this example, we will demonstrate how to change the exposure technique for the selected organ program. The exposure technique can be changed from two point to one point by selecting the correct AEC chamber for the normal patient size. Click the save as default log button. A warning message pops up. Click OK to continue. Another warning message pops up to remind you that all three patient sizes must use the same exposure technique. Select the small and large patients eyes confirmed that the dose and KV values are suitable for the patients eyes. Change the values of necessary. Click the save as default OGP button again to save the changed values into the OGP database. These changed values will then be the default exposure parameters in future exams. The values of KVM as milliseconds dose density measuring field tube load the CPU filter and focus can be changed and saved as default organ program.
SIEMENS ER 202 ER Healthineers Emi Patient I nformation OGP List List Patient ER 2021 03072000001 •skull 143 Other patient Size patient Site size Age: SIEMENS Warning Accession No.; of patient three patient Sites patient all three patient to Device I Device Healthineers Grid Type; Grid Type: Grid Status; Status; Collimator Size; size Size; Collimator Tube Guide