MULTIX Impact: Creating a New OGP Video

This video illustrates how to create a new OGP on the Image Station software.

This video will illustrate how to create a new organ program. Begin by opening service mode, then select the organ program editor button. A password dialog will be displayed. Log in as the manager. Select the organ program tab. Select the body part that you are interested in. In this example, the hand is selected. Click new and then enter and select the required data and parameters in the fields. Note that the ID must be unique for each organ program. Link anywhere from one to 4X parameter templates based on your system configuration and one image processing template to the organ program. Select an acquisition type as a default workplace. Verify that all organ program X Ray parameters and image processing parameters are set as desired. Then select save and select OK when the confirmation dialog appears.

Opefiburation Operation @ Siemens Healthcare GmbH, 2019 Tool Properties Processing Configuration Opefiburation OGP Editor Search Procedure list Procedure OGP OGP list Archiv... Configuration PEPconnect RePconnect RIS Body part Al Hand OGP list Hand PA Body part Snap Shot Statistics Audit Trail Mapping Attribute X-Ray Parameter Image processing parameter Test Image Tube Load ID : Hand Acq type Default Detailt Hand PA NEW Anatomic... Anatomic.. Hand Name : Name : Hand AP Table User setting C-spine AP+LAT+OBL C-spine AP+LAT+OBL C-spine OBL Exp mode : Acq mode * Name Hand PA wall Image flip.. Image lat... Image flap... Image rot... Flavors C-spine flexion General Processing Properties Exam kall Site Info. Setting OProper-ties. Properties Troperties Processing Troperties _Properties Cassette T-spine extension Humerus AP+LAT * Focus . Name Image lat... Description Hand PA Image flip.. Hand PA+OBL C-spine swimmers Image Register Disc Display Config. Troperties Manager Manager. Acq mode Acq type DR RAD Smooth : X Hand PA Image lat... Image rot... 180 270 Exp mode : 2pt * Focus . Hand User High Contrast Hand SID . 115 Image label * Focus . Finely Detailed Tool Passvword Passvwrd Pelvis AP SID . Customized Image label Anatomic... Anatomic.. Grid type Acq type Service View posit... : X Hand PA Cancel Siemens Grid type Remoster Remove Delete Remoster Delete Remove Register DICOM T-spine LAP Not defined Filter type : 0.0mm . 0.0mm Image flip.. Image lat... Image label PMS Target EXI 250 Save Name Assistance Access Filter type : 0.0mm in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the Parameters Collimator lengthr Collimator cengter. Collimator center . : Center Image flip.. Image flat... C-spine AP+LAT SID . 115 View icon Sensitivity Densitivity Manager Save * Name Value Minimum Maximum Medimum Collimator center . Collimator cengthr. Collimator length : 12.21N(31CM) . 12.21N(31CM) Image rot... Image rot.. 180 270 Acq type Grid type Contrast 0.1 0.5 Position Position g.. 250 Procedure information Collimator length Detailt Default 2.6 2.4 1.1 0.1 1.8 0.5 0.7 0.1 8.1 Collimator width : Collimator w'idth : Collimator W'idth : 8.271N(21CM) 8.271N(21CM) Image label Image lat... Filter type . 0.0mm : 0.0mm Demo ima... Demo ima.. ecm . Latitude * Name 2.4 10 90 ecrn . Cancel Collimator... : 8.271N(21CM) kV mAs Tube load Patient size kV mAs Tube load TOD Noise 0.7 authority. Offenders will be Liable for damages. Collimator lengthr Collimator cengter Collimator cengthr Collimator center : Center View posit... Not defined Target EXI Collimator... Yes No Large 10 10 250 WLShift wwshift kV ms m.. AE... Dose Tube... T TOD Maximum Minimum Medimum 1.8 80 Collimator leth Collimator Wdngth . 12.21N(31CM) : 12.21N(31CM) Target EXI 20 250 10 -8 10 Orientatio... Default Description Small wmall ALL names and data of patients, parameters and configuration dependent designations are fictional and examples only. 250 View icon Orientatio... Not defined Enable LifeNet Yes Remove Edit Print New Name Delete Cancel Siemens

  • MULTIX Impact
  • Multix
  • RAD 3
  • Impact
  • Multix
  • Multx
  • Impac
  • MULTIX Impact Creating a New OGP
  • Creating New OGP