MULTIX Impact Smart Virtual Ortho with Camera

This job aid will cover the steps of how to set up and perform Smart Virtual Ortho with the camera on the MULTIX Impact System.

MULTIX Impact Smart Virtual Ortho (With Camera) Siemens Impact VA20 Prerequisites: ✓ Smart Virtual Ortho is configured. ✓ An Ortho Long Leg or Full Spine organ program is selected. ✓ The collimator must be in the 0° lock-in-position. ✓ The tube unit is positioned in detent and centered to the detector at the correct SID. ✓ For Ortho acquisitions at the wall stand, the detector unit must be in the 0° position. ✓ Position the patient upright on the Ortho Stand or flat on the table-top. ✓ The Ortho menu is displayed on the Tube TUI. Steps: 1. Measure the Table-to-Object Distance (TOD) and enter on the Control Console 2. For Region of Interest Auto-Detection, select the camera icon on the imaging system. The ROI is automatically detected and shown with a blue frame on the imaging system. lealthineers 1 1 ............... ? Patient Information Jane Doe 202009180953251 Other 40Y 9-18-1979 202009180953251 S Device Information Und Tyaw: Ne Grid Golf Statut Geld OUT CHANCE Gold Orientation: Grid_Vertical VD 180cm 43.0008 -. 35.0on End of bom Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 09/15/2021 | HOOD05162003220302 Healthineers Job Aid: Smart Virtual Ortho (with Camera) Effective Date: 09/15/2021 | HOOD05162003220302 3. If necessary, adjust the upper and lower border lines to define imaging range on the Control Console 4. Collimating the left or right border lines will adjust the width of your region of interest (ROI). Performing Acquisitions and Post-Processing: 1. Ensure the correct exposure factors are set for each required step in the ortho series at the Control Console. 2. With the patient maintaining positioning, press and hold the acquisition switch until all exposures are completed and tube returns to it’s original position. • Note: Releasing acquisition switch before all images are acquired will result in the interruption of the series. Exam will need to be repeated in its entirety. 3. Once series acquisition is complete, release the switch. Auto composing of the image then occurs. Name: jone, , jone Institution nan Sex: M Aquisition date: 2020-11-12 View Composing ID: 202011121356111 Acquisition time: 14:30:29 Tool Date of birth: 1992-06-16 Operator: Mgr Age: 028Y Request ID: Accession no.: 2020111213551 Measurement L mAS: 1.01 kV: Reference EXI: 250.00 Current EXI: 269.18 DI: 0.32 Matrix size: 3033x6832 Filter type: NONE OOP name: Ortho_Finished WC: 1872 Grid type: NONE manual adj ents performed WW. 3136 4. Measurements can be made to the composed image prior to saving under the “View” tab. 5. Adjustments to the composing process can be made in the “Composing” tab. Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 2 of 5 Job Aid: Smart Virtual Ortho (with Camera) Effective Date: 09/15/2021 | HOOD05162003220302 View Composing Composing controls Steps 5 Transition mode Blend Transparent 6. Steps are adjusted from first to last by shifting the lower image left, right, up, or down relative to the image above. Select “Next” after each image . 7. Select “Save” once completed to finish the process and save the composed image. Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 3 of 5 Job Aid: Smart Virtual Ortho (with Camera) Effective Date: 09/15/2021 | HOOD05162003220302 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter co llectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written auth ority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 4 of 5 Job Aid: Smart Virtual Ortho (with Camera) Effective Date: 09/15/2021 | HOOD05162003220302 Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Healthineers Page 5 of 5