Multix M: Detector Calibration

Multix M: Detector Calibration

Multix M: Detector Calibration 1. Perform a Calibration annually or if issues with image quality arise. The system should be ON for 30 minutes prior to performing a Calibration. 2. Place the detector on the floor. Set the SID to 200cm. Open the collimator to illuminate the entire surface of the detector. 3. Select SYSTEM>CALIBRATION. CALIBRATION 4. The system will display the date of the last calibration. If not, a warning will appear instructing you to perform one. Last Calibration 20/01/2004 15:52 5. Select START. Start 6. Follow the instructions in the upper left corner of the screen. Four exposures should be performed Preparing as instructed. Wait a moment. 7. Calibration result will appear. Calibration has succeeded. 8. Select FINISH. Multix M: Detector Calibration SIEMENS www.usa.siemens.com/healthcare 388 Effective Date: 09/08/2016