Multix M: Detector Self Test

Multix M: Detector Self Test 

Multix M: Detector Self-Test 1. Perform a Self-Test once every month or if issues with image quality arise. 2. Place the detector on the floor. Set the SID to 200cm. Open the collimator to illuminate the entire surface of the detector. Ensure the detector is free from artifacts, i.e., lead markers. 3. Select SYSTEM>SELF-TEST. TEST1 Passed. TEST2 Passed. SELF TEST TEST3 Passed. TEST4 Passed. TEST5 Passed. 4. Select START TEST. Start Test 5. Follow the instructions in the upper left corner of the screen. One exposure should be performed as instructed. 6. Five tests will be performed. If a Calibration has not been performed first, test one will fail. 7. Result of Self-Test will appear. System 8. Select EXIT. Multix M: Detector Self-Test SIEMENS www.usa.siemens.com/healthcare 389 Effective Date: 09/08/2016