Performing Autocheck on the Atellica® CH Analyzer Job Aid Version 1.25 and Below
This job aid provides instructions on how to perform the as needed maintenance task of Autocheck for the Atellica® CH Analyzer.
Atellica® CH Analyzer SIEMENS Performing CH Autocheck Job Aid Healthineers 1. Ensure analyzer is in Diagnostic State: SIEMENS Atellica® Solution Version 1.14.0.2974006 P-P1-03 A. Select System at the Command bar A System Samples K Inventor Calibration Qc Patient Orders Worklist Maintenance Setup Status Logs B Operator Diagnostics About B. Select Operator Diagnostics C-P1-06 C C-P1-06 P1-060 C. Select the CH Icon Subsystems Diagnostic Name Description Lamp Autocheck 4 CH Autocheck 5 Perform D. Select Enter Diagnostics > Yes Cuvette 2 IMT 2. Wait for analyzer to enter Diagnostic State Probes Fluid 3. Under Subsystems, select Autocheck 3 Autocheck 4. Under Diagnostic Name, select Autocheck Mixer Inspect and Clean 5. Select Perform > Yes 6. Wait for Autocheck to complete which takes approximately 11 minutes Select Exit Diagnostics > Yes D Enter Diagnostics 7. Ext Diagnostics ? PEPE ] CSE OS Monday, June 04, 2018 2:30 PM 1 Template Effective Date: 05/27/2019 | HOOD05162003010284 Effective Date: 24-SEP-2020 | HOOD05162003126734 Unrestricted © Siemens Healthcare GmbH, 2020 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2020 Global Siemens Healthineers Headquarters Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84 0 siemens.com/healthineers 2 Template Effective Date: 05/27/2019 | HOOD05162003010284 Effective Date: 24-SEP-2020 | HOOD05162003126734 Unrestricted © Siemens Healthcare GmbH, 2020
- attelica
- attellica
- atelica