Performing Automated Atellica® Sample Handler Autocheck Job Aid Version 1.25 and Below

This job aid provides instructions on how to perform the daily maintenance task of Automated SH Autocheck for the Atellica® Sample Handler.

Atellica® Sample Handler SIEMENS Performing Automated SH Autocheck Job Aid Healthineers SIEMENS Build 1.7.0 DL-P1-07 The SH Autocheck is an automated maintenance activity 1 System Samples Inventory Calibration QC Patient Orders Worklist Maintenance Setup that tests the health of various SH components. 2 Schedule Report Log Diagnostics Data Restore Software Updates Copy Configuration Filter Module: All Frequency: All Time: Next 24 hours Performed By: All The system alerts the operator 15 minutes before the Module All Module Activity Scheduled Sun-28 Mon-29 Tue-30 Wed-31 Thu-01 Fri-02 Sat-03 4 Perform ... scheduled activity begins. After the system initiates the SH Frequency All CH GGCKKGSCI-CH IMT Daily Cleaning :: 05/31/2017 2:00 V V V Mark as Completed Time Next 24 hours CH Autocheck, the SH pauses sample processing until the GGCKKGSCI-CH CH Daily Maintenance : 05/31/2017 4:00 V V Not Perform .. Performed By All IM GGCKKGSCI-IM IM Autocheck 05/31/2017 3:30 autocheck completes in approximately 5 minutes. V V V V Search Text Enter the activity name. SH GGCKKGSCI-SH SH Autocheck 3 Procedure Details .:: 05/31/2017 1:00 V V The task will run automatically daily, but can be run Scheduled by Operator VMM VMM Atellica Magline Transport Autocheck :!: 05/31/2017 1:00 V V Edit/View ... Scheduled Automatically Database Backup :: 06/01/2017: 0:30 V V Delete manually if needed: Scheduled by Service IM GGCKKGSCI-IM Cleaning the Water Container 06/01/2017 1:00 V Failed 1. On the Command bar, select Maintenance Overdue GGCKKGSCI-IM Inspecting and Emptying the IM Water Trap :06/01/2017 1:00 V V Inprocess Database: Optimizing .:: 06/01/2017 3:00 V 2. Select the Schedule tab Completed IM GGCKKGSCI-IM IM Weekly Maintenance :06/01/2017 5:00 V V x Not Performed Database: Archiving and Deleting .:: 06/01/2017 7:00 -- -- 3. Select SH Autocheck With Comments IM Www Activites GGCKKGSCI-IM Cleaning the Exterior of the Aspirate Probes 06/01/2017 7:00 IM 4. Select Perform GGCKKGSCI-IM Cleaning the Reagent Probe Rinse Stations 06/01/2017 8:00 Add Activity .. Set Warning. 5. Wait for autocheck to complete ? B) LabManager Monday, April 17, 2017 10:02 AM HOOD05162002822815 © Siemens AG 2013. All rights reserved. Effective Date: 3/12/2018 Page 1