PET image Sequence Order - Biograph mCT and Horizon
This is a step by step to achieve all PET images being displayed in the same direction (craniocaudal or caudocrainial), the PET scan direction and patient position must be set up correctly on Biograph mCT VG60 and Horizon VB10.
PET Sequence Order Biograph mCT and Horizon Description: This is a step by step to achieve all PET images being displayed in the same direc- tion (craniocaudal or caudocrainial), the PET scan direction and patient position must be set up correctly on Biograph mCT VG60 and Horizon VB10. Not applicable for VG70 or VJ20. For Biograph mCT VG60 and and Biograph Horizon VB10 1. When selecting the protocol in the patient model dialog, you must select the correct patient orientation. Patient Model Dialog @ Adult al Energy Protocol o Replace @ Child Append Topogram c Cut SCU o Keep RT Auto reference lines None ecials PET API Language English OK Cancel 2. In the PET chronical on the Routine subtask card, the PET Scan direction must correctly display the direction of the PET sequence order desired both for single bed or multibed aqcuistions. SIEMENS siemens.com/healthineers Healthineers . PET Sequence Order – Biograph mCT and Horizon Effective Date: 03/13/2018 | Quality HOOD05162002829091 PETCT_WholeBody (Adult) Patient, My 787.7 mm 24 min 10 mCi Topogram Isotope -18 Pharm Fluorodeoxyglucose CT WB Dose 10.00 mCi Pause Date 26-Oct-2012 Time 15:12:18 PET WB Scan range Match CT Range 2 No. of beds Scan duration : 3 min Range: Begin End Table: Position Height Scan direction Load Hold Recor -1454.0 -666.0 -600.0 100.0 Caudocranial Recon Routine Scan Recon Auto Tasking a. The CT sequence is not influenced by the CT scan direction. It is influenced by the CT recon direction. b. For best practices have both CT recon direction and PET scan direction match to have both CT and PET sequence orders match. Biograph mCT/Vision VG70 and Biograph Horizon VJ20 1. On the PET chronicle on the Autotasking subtask card, the PET image Order can be set in the desired direction regardless of PET scan direction. Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this Siemens Healthineers Services Page 2 of 3 training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018
- PET image Sequence Order - Biograph mCT and Horizon
- craniocaudal or caudocrainial
- BiographmCTVG6 and HorizonVB10