Reinstalling the Automatic QC Cartridge Video

This video provides instructions on how to properly reinstall the Automatic QC Cartridge on your RAPIDPoint® 500 System, including how to check the black valve lever for proper alignment prior to reinstalling the cartridge.

Open the cartridge connector by pushing in and then sliding the connector to the right. Wait for the cartridge to release. Remove the cartridge. If you're reinstalling the same automatic QC cartridge, turn the cartridge around to check that the black valve lever is properly aligned between the arrows. Insert the cartridge in the system, then push firmly on the raised dot until you hear the cartridge lock in place. Slide the cartridge connector to the left to close it. The installation is complete then touched the continue button.

AutomaticOC