RAPIDPoint® 500e Blood Gas System Calibration Job Aid
This job aid provides instruction on how to perform calibration.
Calibration Job Aid RAPIDPoint® 500e Blood Gas System Calibration Overview The system automatically performs calibrations at prescribed intervals: · 1 point calibrations are scheduled to occur regularly at 30 minute intervals. It adjusts either the offset or the slope drift for a parameter · Every fourth scheduled calibration is a 2-point calibration (every 2 hours). It adjusts both the offset and the slope drift for a parameter and the zero for tHb · Every fourth 2-point calibration is a full calibration (every 8 hours). It adjusts both the offset and the slope drift for a parameter, and the zero and slope for tHb · During calibration, if the system detects a problem for a parameter, the system repeats the calibration for as many as two times · If the calibrations are not successful, the system turns the parameter off. The parameter displays with either a line or an “X” Indicates that this parameter is not available because the sensor has failed calibration Indicates that this parameter has failed calibration and is unlikely to pass further calibration Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/11/2019 | HOOD05162002998095 Healthineers Job Aid Effective Date: 18/Mar/2020 | HOOD05162003075280 Job Aid | Calibration Effective Date: 18/Mar/2020 | HOOD05162003075280 If a parameter is turned off, you may perform a manual calibration. 1. If prompted, enter your password, or your password and operator ID 2. At the Analysis screen, select the System button Ready 13:56 ? 07/01/2019 Glu 2 a PH DC PCO2 K+ Lac OC pO2 Ca++ tHb CI- nBili (Optional) Start Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 2 of 6 Job Aid | Calibration Effective Date: 18/Mar/2020 | HOOD05162003075280 3. Select Calibrate Status 10:49 07/02/2019 Cartridge Status W QC Replace Analyzed 10 10 Days Remaining 22 4 27 Replace Port Exp. Date 07/25 07/07 07/30 Exp. Time 08:54 08:54 10:40 Events Log Shutdown Clean Screen Calibrate System Info ... Setup 4. Select the calibration type and select Start. Some calibration types may not be available because an automatic calibration is scheduled to begin shortly Calibrating 09 52 07/02/2019 Calibration Type 1-point 2-point Full Start Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 3 of 6 Job Aid | Calibration Effective Date: 18/Mar/2020 | HOOD05162003075280 5. The calibration begins. If you want to interrupt the calibration to analyze a priority sample, select STAT Calibrating 09:53 07/02/2019 V a nly CAUTION: The system is busy. Do not insert a sample device. The sample may be lost or contaminated with reagent. Touch STAT to analyze a sample. p 1:34 STAT Start Glu Lac 6. When the calibration is finished, the Analysis screen appears Ready ? 14 00 07/01/2019 a Na+ Glu OC) PH pCO2 K+ Lac pO2 Ca++ tHb CI- nBili (Optional) a Start Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 4 of 6 Job Aid | Calibration Effective Date: 18/Mar/2020 | HOOD05162003075280 Calibration Report Formats Depending on the options selected in Setup, the system prints a full calibration report or a system status report. The status report describes the calibration results as passed or failed and identifies any failed parameters. Examples of the status report are below. RAPIDPoint® 500e RAPIDPoint® 500e System Status System Status 07/02/2019 13:30 07/02/2019 13:34 System ID 0500-33788 System ID 0500-33788 Na+ Failed Others passed Sensors passed D33 Slope Error Na+ Additonal cal required Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 5 of 6 Job Aid | Calibration Effective Date: 18/Mar/2020 | HOOD05162003075280 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. RAPIDPoint is a registered trademark of Siemens Healthcare Diagnostics. Copyright © Siemens Healthcare GmbH 2020 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 6 of 6
- Rapid point
- Rapid Comm
- blood gas
- arterial
- capillary
- calibrate