RAPIDPoint® 500e Blood Gas System Cleaning and Disinfecting the System Job Aid

This job aid provides instruction on how to clean and disinfect the system.

Cleaning and Disinfecting the System RAPIDPoint® 500e Blood Gas System Refer to the Operator’s Guide when performing the following procedures. Cleaning the Touch Screen Clean the touch screen as needed to remove dust, dirt, or splatters from the screen and disinfect the screen surface. Supplies needed: lint-free cloths moistened with 0.5% sodium hypochlorite solution so that it is wet but not dripping 1. If prompted, enter your password, or your password and operator ID 2. Select the System button and then select Clean Screen Status 13:43 ? 07/01/2019 Cartridge Status 2 M da W Replace Analyzed 4 Days Remaining 23 5 Replace Port Exp. Date 07/25 07/07 Exp. Time 08:54 08:54 Events Log Shutdown Clean Screen Calibrate System Info ... Setup Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/11/2019 | HOOD05162002998095 Healthineers Job Aid Effective Date: 30/Mar/2020 | HOOD05162003076203 Job Aid | Cleaning and Disinfecting the System Effective Date: 30/Mar/2020 | HOOD05162003076203 3. While the Clean screen displays, wipe the screen with the wet cloth. After 20 seconds, the system returns to the System screen Status 09 31 07/02/2019 Clean the screen surface. :16 4. Wait 10 minutes after applying the sodium hypochlorite solution then access the Clean screen again and dry. After 20 seconds, the system returns to the System screen 5. Select the Continue button on the lower right to return to the Analysis screen Status 13:43 07/01/2019 Cartridge Status Replace Analyzed 4 4 Days Remaining 23 5 Replace Port Exp. Date 07/25 07/07 Exp. Time 08:54 08:54 Events Log Shutdown Clean Screen Calibrate System Info ... Setup Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 2 of 4 Job Aid | Cleaning and Disinfecting the System Effective Date: 30/Mar/2020 | HOOD05162003076203 Cleaning and Disinfecting the Exterior Surfaces Clean the exterior surfaces as needed to remove dust, dirt, and splatters from the surfaces, and disinfect the surfaces. Supplies Needed: lint-free cloths, 0.5% sodium hypochlorite solution Please follow these precautions when using cleaning agents: · Do not use any solution containing benzalkonium chloride or other quaternary ammonium compounds · Do not wet the sample port or the sensor contacts for the measurement and AQC cartridges · Do not spray cleaning solution or other fluids on or into the sample port or the area behind the measurement and AQC cartridge when cleaning surfaces · Do not spray cleaning solutions or other fluids on the optics head assembly which is located behind the measurement cartridge · If you choose to use an alternative to bleach, consider the disinfectant wipes which Siemens has evaluated and approved. Refer to the Operator’s Guide for a list of products Procedure: 1. Moisten the cloth with the sodium hypochlorite solution so that the cloth is wet but not dripping 2. Wipe the exterior surfaces of the system with the wet cloth 3. Wait 10 minutes after applying the sodium hypochlorite solution and then dry the surface Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 3 of 4 Job Aid | Cleaning and Disinfecting the System Effective Date: 30/Mar/2020 | HOOD05162003076203 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. RAPIDPoint is a registered trademark of Siemens Healthcare Diagnostics. Copyright © Siemens Healthcare GmbH 2020 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 4 of 4

  • Rapid point
  • Rapid Comm
  • blood gas
  • instrument
  • surface