RAPIDPoint® 500e Blood Gas System Removing and Reinstalling an AQC Cartridge
This video provides instructions on how to perform the task of removing and reinstalling an AQC Cartridge on the RAPIDPoint® 500e Blood Gas System.
View this video before you begin to replace the automatic QC cartridge. Select system button. Select automatic cartridge button. Select replace button. Select yes to replace the automatic cartridge. Open the automatic QC cartridge connector by pushing in and then sliding the connector to the right. Wait for the cartridge to release. Remove the cartridge. If you're reinstalling the same automatic cartridge, turn the cartridge around to check that the black valve lever is properly aligned between the arrows. Reinsert the automatic cartridge into the system, then push firmly on the Rays dot until you hear the cartridge lock in place. Slide the automatic QC cartridge connector to the left to close it. The wait screen appears while the system prepares the cartridge. The analysis screen appears when complete. The installation is complete. Play the video again if you need help replacing the automatic cartridge.
@ 2019 Siemens Healthcare Diagnostics Inc. All rights reserved. RAPIDPoint and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc., or its affiliates. 12:14 1? 1b 13 Ready me ddy xeaay Kea dy status bra tus >tatus 8eady yea ay 12/147018 All other trademarks are the property of their respective owners. Please note that the learning material is for training purposes only! Cartrx±ge Status Cartr•dge Status Glu Na• Replace Date For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively Automatic QC SIEMENS . SIEMENS The AutomatKQC cartridge does not need replacing. Aru The AutomaticQC cartridge does not need replacing. Are "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means pco. pcm Lac Analyt Analyzed The system is temporarily busy. Please wait. you sure you want to replace the AutomaticQC cartridge? Replace Port Days Remaiming Days Remaimung Days Rerna.n.ng 13 27 substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily Touch Yes to replace the AutomaticQC cartridge. Automatic QC Healthineers Exp. Time Exp, Date Date 12/24 12:14 01/11 28 Touch No to return to the Status screen. tHb reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as Exp. rime 10:35 Shutdown Events Log Events your main reference, in particular for relevant safety information like warnings and cautions. Healthineers cr (Optional) nBili Note: Some functions shown in this material are optional and might not be part of your system. Automatic QC Clean Screen Automatic QC Screen SIEMENS • 2:56 Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially Effective Date January 7, 2019 Calibrate available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not HILSlHOOD05162002952940 Yes guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. SIEMENS Copyright @ Siemens Healthcare GmbH 2019 System System Info.. Setup Start Startus The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Global Siemens Healthineers Headquarters Offenders will be liable for damages. Henkestr. 127 91 052 Erlangen, Germany All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Phone: +49 9131 84 All rights, including rights created by patent grant or registration of a utility model or design, are reserved. siemens.comlhealthineers
- Rapid point
- Rapidpoint
- Rapidpoint 500e
- Rapidpoint500e
- 500e