RAPIDPoint® 500e Blood Gas System Replacing a Wash Waste Cartridge without AQC
This video provides instructions on how to perform the task of replacing a wash waste cartridge without ACQ on the RAPIDPoint® 500e Blood Gas System.
View this video before you begin to change the wash waste cartridge. Make sure you have a new wash waste cartridge. Select system button. Select wash waste cartridge button. Select replace button. Select yes to replace the wash waste cartridge. Open the door. Remove the wash waste cartridge. Get a new wash waste cartridge. Insert the cartridge in the system, then push firmly on the dot until the cartridge locks in place. Close the door. The wait screen appears while the system prepares the cartridge. The analysis screen appears when the system is ready. The installation is complete. Play the video again if you need help replacing the wash waste cartridge.
@ 2019 Siemens Healthcare Diagnostics Inc. All rights reserved. Start RAPIDPoint and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc., or its affiliates. 14 12 30 12 JO 12:31 1? 31 14 28 Ready 14 14:28 00B Seatup Startus 12/142018 All other trademarks are the property of their respective owners. Start Please note that the learning material is for training purposes only! Cartridge Status RAPIDPoint• 405/500 Systems Glu pH 405/500 Systems Glu pH 500 systems RAPIDPoint 405/500 systems For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively Replace Wash/Waste Cartridge SIEMENS Lac "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means he Wash/Waste cartridge does not need replacing. 2806701219 pcoz pco: Lac Analyzed The system is temporarily busy. Please wait. Are you sure you want to replace the Wash/Waste Replace Port substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily Days Remaining 27 cartridge? 10491449 Exp. Date 01/11 12/24 pcoa tHb ouch Yes to replace the Wash/Waste cartridge. reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as Exp. Time Touch No to return to the Status screen. 14:28 14:22 Shutdown your main reference, in particular for relevant safety information like warnings and cautions. RAPIDPoint• 405/500 Systems Events Events Log 405/500 Systems (Optional) Healthineers ' Cl- Cl- nBili nBiIi Wash/Waste Cartridge RAPIDPoint 500 Systems 500 systems Note: Some functions shown in this material are optional and might not be part of your system. 405/500 Systems Clean Screen 500 Systems Systems Wash/Waste Cartridge stems Cartridge Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially 405/500 Systems Effective Date January 7, 2019 500 systems RAPIDPoint 405/500 Systems Calibrate Wash/Waste Cartridge 2806701219 Cartridge available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not cr F02Hb. cr. 08610051 FCOHb, pH, pcm. Na•. K•, ca••. pH. p02. Na•, K•, REF Yes cr, F02Hb. No guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. cr. Lac. FCOHb. S/EMEWS • FCOHb, FMetHb, FHHb pH, pC02. p02. Na•. pH. p02, Na•. ca••. pH, pm, Na•, 08610051 pH, pc02. Na•. Copyright @ Siemens Healthcare GmbH 2019 cr. Lac. REF System Info... cr. F02Hb, Setup FCOHb, FMetHb, FHHb The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. 2018-10-18 Start S/EMEWS • 2018-10-18 Global Siemens Healthineers Headquarters Offenders will be liable for damages. Healthineers 2018-10-18 SIEMENS SIEMENS 12 Henkestr. 127 2018-10-18 S/EMEWS • SIEMENS 91052 Erlangen, Germany All names and data of patients, parameters and configuration dependent designations are fictional and examples only. 14 SIEMENS Phone: +49 9131 84 14 All rights, including rights created by patent grant or registration of a utility model or design, are reserved. siemens.comlhealthineers Healthineers
- Rapid point
- Rapidpoint
- Rapidpoint 500e
- Rapidpoint500e
- 500e