RAPIDPoint® 500e Blood Gas System Replacing an AQC Cartridge
This video provides instructions on how to perform the task of replacing an AQC cartridge on the RAPIDPoint® 500e Blood Gas System.
View this video before you begin to replace the automatic QC cartridge. Make sure you have a new automatic cartridge. Select system button. Select automatic cartridge button. Select replace button. Select yes to replace the automatic QC cartridge. Open the automatic QC cartridge connector by pushing in and then sliding the connector to the right. Wait for the cartridge to release. Remove the cartridge. Get the new automatic cartridge. Remove the yellow card from under the lever. Then press down firmly near the raised dots to close and lock the lever in the cartridge. Be sure the lever locks in place. Insert the cartridge into the system. Take care to ensure that the cartridge slides into the bracket located toward the back of the system. Then push firmly on the raised dot until you hear the cartridge lock in place. Slide the automatic cartridge connector to the left to close it. The wait screen appears while the system prepares the cartridge. The analysis screen appears when complete. The installation is complete. Play the video again if you need help replacing the automatic QC cartridge.
@ 2019 Siemens Healthcare Diagnostics Inc. All rights reserved. RAPIDPoint and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc., or its affiliates. *eaay 1? 1b 13 13 status Kea ay Neaay 12:14 12/147018 All other trademarks are the property of their respective owners. Please note that the learning material is for training purposes only! Cartr•djge Status Cartr•dge Status status Startus Na• Glu Replace For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively Automatic QC SIEMENS . the AutomatjcQC cartridge does not need replacinq. Are The AutomaticQC cartridge does not need replacing. Are "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means pcm SIEMENS • Analyzed The system is temporarily busy. Please wait. Touch Yes to replace the AutomaticQC cartridge. you sure you want to replace the AutomaticQC cartridge? Healthineers Replace Port Days Remnajntng 13 Days Remajntng substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily Touch Yes to replace the AutomaticQC cartridge. Automatic QC Date 12/28 01/11 12/28 Touch No to return to the Status screen. tHb reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as 10:35 12:14 Shutdown Events Events Log your main reference, in particular for relevant safety information like warnings and cautions. Healthineers (Optional) Healthineers - nBili Automatic QC Note: Some functions shown in this material are optional and might not be part of your system. Clean Screen Screen Automatic QC 2:57 Automatic QC Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially Effective Date January 7, 2019 Calibrate available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not Yes guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. RAPIDpoint sooe Copyright @ Siemens Healthcare GmbH 2019 System Info... System Info.. sooe Setup Start The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Global Siemens Healthineers Headquarters Offenders will be liable for damages. Henkestr. 127 91 052 Erlangen, Germany All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Phone: +49 9131 84 Automatic QC All rights, including rights created by patent grant or registration of a utility model or design, are reserved. siemens.comlhealthineers sooe
- Rapid point
- Rapidpoint
- Rapidpoint 500e
- Rapidpoint500e
- 500e