RAX: Reducing OID on Table Exams

This job aid covers steps to follow to help reduce OID

RAX: Reducing OID on Table Exams XPRF To help reduce magnification on table (T) exams (such as Pelvis and Abdomen studies), it may be necessary to reduce the OID by moving the detector closer to the underside of the table. After positioning the patient as you normally would for the exam, use the Manual Up/Down movement button on the detector to move the detector closer to the table. NOTE: Manually moving the detector close to the table brings it into the collision zone of the table. Tube/Detector tracking may not be possible while the detector is in this position. Once the image is taken, it may be necessary to manually move the detector down out of the collision zone before proceeding to the next image. Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 01/21/2021 | HOOD05162003162827 Healthineers Job Aid | RAX: Reducing OID Effective Date: 01/21/2021 | HOOD05162003162827 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter co llectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written auth ority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Healthineers Page 2 of 2