Recovering from the Sample Handler Robot Dropping a Sample Container Video Version 1.25 and Below
This video will explain how to recover from the robot dropping a sample container on the Atellica Sample Handler.
Welcome to the Atellica solution. Recovering from the sample handler robot. Dropping a sample container video. To recover from the Atellica sample handler robot, dropping a sample container first determine which sample handler and or atellica magline transport covers need to be removed to access the drop sample container. Then stop the sample handler and then the atellica magline transport and unlock the covers as applicable to access the dropped sample container to stop the sample handler, first select the system button on the command bar. Next, select the Status tab and then the sample handler module icon on the sample handler status screen. Select the stop button. Select the buttons to unlock the sample handler. Front and back covers as appropriate. If needed, also stop the atellica magline transport by selecting the Atellica magline transport to have. And then selecting the stop button. Left open the sample handler. Front and back covers if applicable. Retrieve the drop sample container and then if needed to clean this bill. Remove the sample racks from the sample drawers. Clean this build by first wiping the area with lint, free tissues or gauze saturated with a 10% bleach solution, then wiping the area with lint free tissues or gauze saturated with water and finally drying the area with clean lint, free tissues or gauze. Inspect the sample handler robot gripper orings. If any orings show signs of wear or damage, replace all four rings. Close any open sample handler covers and lock. The sample handler covers, as applicable by selecting the lock back cover and or lock front cover buttons on the sample handler status screen. When the local Technical Support provider informs that it is safe to do so, start the SH by selecting the resume button on the sample handler system status screen. If the drop tube was retrieved through the back cover, confirm the calculus. The storage area contents by selecting the confirm calculus. See storage content button on the sample handler system status screen. If the contents of the calculus, the storage area were not changed, select the operator did not change the contents, and then select OK. If applicable, start the atellica magline transport by selected the resume button on the Atellica magline Transport System status screen. This completes the procedure.
Atellica• Solubion Atellica• Solubon AtelLica@ Solution AteUlica• Solution AteUica• Solution S0[ution AtelLica@ Solution Disclaimer О Мет: С Мег: Orders ссос Please note that the learning material is for training purposes only! IA.03 For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively Recovering from the "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. Sample Handler Robot Capabilities The Operator's Manual shall be used as your main reference. in particular for relevant safety information like warnings and cautions. p•peuuoo Dropping a Sample Robot Note: Some functions shown in this material are optional and might not be part of your system. А*сд rcs Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in Container CaBQC your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. Connected to PCC РСС The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will s•gwqedeo be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Confirm C*.QC Confirm Сов All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright @ Siemens Healthcare GmbH 2020 соме. Cover Cover Siemens Healthineers Headquarters\Siemens Healthcare GmbH\Henkestr. 1271 91052 Erlangen, GermanyiTelephone: +49 9131 40 40 28 гоп дм Master Template HOOD05162T03052540 1 Effectve Date: 26-Nov-2019 Master Template HOOD05162C03052540 1 Date: 26-Nov-2019 uoqnps
- Atelica
- Attellica
- Attelica