Removing Tubes from the SH Robot Grippers Manually Video Version 1.25 and Below
This video will explain how to remove tubes from the robot grippers on the Atellica Sample Handler.
This procedure should only be performed if the system displays an alert. To manually remove a tube in the sample handler. Robot gripper. When this message displays, the system will automatically be put in the stopped state. To remove tubes from the robot gripper manually, first open the sample handler front cover by selecting the system button on the command bar. Next, select the Status tab and then the sample Handler tab on the sample handler status screen, select front cover. Wait for the lock icon to appear unlocked and open the front cover within 30 seconds. Manually move the sample handler robot to the front of the sample handler. Do not hold the SH robot grippers to move the robot. Only move the robot by holding the robot body. If the robot is out of reach from the front of the sample handler, open the sample handler back cover to move the robot into position while holding both the tube and robot gripper. Turn the robot gripper manual release counterclockwise to release the tube and place the tube in Iraq. While still holding the gripper, visually inspect that the four O rings are in place and not dislodged. Close the sample handler front cover if appropriate. Also closed the sample handler back cover. Lock the SH front cover by selecting the lock front cover button on the sample handler status screen. If appropriate, also lock the sample handler back cover to complete the procedure. Start the sample handler by selecting the resume icon on the sample handler system status screen.
SIEMENS Ate I licaTU Solution Ate Healthineers Heotthineers Samples Inventory Cali txabon P atieN O rckrs Setup Please note that the learning material is for training purposes only! CC,'OL operator operator Operator Diagnostics Diagnostics For the proper use Of the software, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") Diagnostics issued by Siemens Healthineers. This material is to be used training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software System VMM .01 CH IA -03 IA-03 and hardware available at the time Of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Processing Remouing mUbes from Remouing TUbes from Note: Some functions shown in this material are optional and might not be part of your system. Tirne Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Capabilities Due to regulatory requirements, the future availability Of said Functionalities in any specific country is not guaranteed. Please contact your local SIEMENS Siemens Healthineers sales representative for the most current information. Robot SH Robot Gripper ManuaLLg The reproduction, transmission or distribution Of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. Tcs TCS All names and data Of patients, parameters and configuration dependent designations are fictional and examples only. cal-QC Heotthineers Cal-QC storage All rights, including rights created by patent grant or registration of a utility model or design, are reserved. connected to pcc Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. DrawM 2 Drawer 2 product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. DrawM 2 Copyright O Siemens Healthcare GmbH 2020 Drawer 4 Drawer I t_Jnkxk Frcnt Cover t_Jncxk Frcnt Covet tncxk Frcnt Covet Front Cover Global Siemens Headquarters Global Siemens Healthineers Headquarters Front Cover Front Cover Henkestr. 127 Back Cover Back Cover Unlock Front Cover UNock Back Cover Cover 91052 Erlangen, Germany Back Cover Phone: +49 913184 0 siemens.com/healthineers
- atelica
- attellica
- attelica