Replacing Dilution Cuvettes on the Atellica® CH Analyzer Video Version 1.25 and Below

This video provides instructions on how to perform the as needed maintenance task of replacing the dilution cuvettes on the Atellica® CH Analyzer.

This video will show you how to change the dilution queue vets on the Atellica Chemistry analyzer. The analyzer alerts for the operator to replace the dilution cuvette segments every four months. All of the dilution Cuvette segment should be replaced at the same time. To change the dilution kovetz the system must be set to diagnostic state. This can be done by first selecting system in the command bar. Next, select operator diagnostics. Select the chemistry module icon. Select enter diagnostics. Select yes when you have entered into diagnostics. Select cuvet in subsystems. Select replacing the dilution ring cuvette segments. Select perform. Select yes, wait for the analyzer to prepare for maintenance. The covers automatically unlock. Open the back cover of the analyzer to the top position. Rotate the dilution ring to access each dilution Cuvette segment to unlock a segment, turn the thumbscrews 1/4 of a turn counterclockwise so that the segment can be lifted out. Lift up to remove the segment and discard the cuvet as per laboratory protocol. Place new segment in the dilution ring. Align the pins to the holes in the segment. Lock the segment in by turning the thumbscrews quarter of a clockwise turn until the screws lock into place. Repeat these steps to replace all the cuvet segments. Close the back cover. To exit diagnostics state, first select continue in the software. Next select close. Select exit diagnostics. Finally, select yes to exit diagnostics. To finish the procedure, an mark is completed, select the maintenance tab in the command bar. Select schedule. Select the chemistry module in the dropdown menu. Select replacing the dilution ring cube at segments. Select the Mark is completed button on the right hand side of the screen. To complete the procedure, perform cleaning the Ch vets. Note that if you are replacing both the reaction Cuvette San dilution cubits at the same time, you need not perform cleaning the Ch vets because performing Ch weekly maintenance is a required step for the replacing reaction cuvettes procedure and Ch weekly maintenance included cleaning the queue vets.

SIEMENS Atellica• Solution Version 1.14.02951003 Atetlica• Solution Version 1.14.02983003 Solution V 9b0 1.14.0.2 3003 Solution V 9b0 t14.02 3003 Atellica• Solution Version 1.14.02983003 Solution version 1.14.0.2983003 SP00107 Atellica Solution Version 1.14.0.2983003 Heotthineers Heoltnineers Heottnineers H 0れin′、 SIEMENS は」 Healthineers Heoltnineers Invert ory Inventory Invent ory Cⅸ P 0当、 Operator Diagnostics Software u , Software Updates Copyconfiguration Copy C Copy Configuration CH ATELLICA•CH ATELLICA-CH ATELLICAVCH ATELLICA ATELLICACH ATELLICA.CH Module: Performed Module: Performed By. Mod、ATEは℃CH eqしenc Trne p System MO物ATELLI'CA-CH ATELUICA-CH ATELLICAVCH MO物ATELLI'CA•CH M 0 Ac Fri•02 S 93 sat-03 S0094 M0m05 100・06 Wed-07 Wed•07 Module Subsystems Diagnostic Name Diagnostic N a me Ooscnption Description Doscnption Name scrip ton ATELLICA-CH 1日℃A H ( 01 0 800 ) 0 01 0 80 ( 0 O n f i CH Reactgm 'Ong Cuvette Sequents Re$*acir" CH Reaction Rim Cuvene Segrnents Resiacir" CH Reacbon Rim Cuvette Segnents the CH Reaction Cuvette Segrnents the CH ReactOn Rim the CH Reacton Rim CH CH Reaction Rim Cuvette Segrnents CH Reacbon Cuvette Segrnents CH Reacbon Rim Cuvette CH Reactan Cuvette CH Reaction Rim Cuvette CH Reaction Cuvene Segrnents Resiacir" CH Reacbon Rim Cuvette Segnænts Reøac•m me CH Reactkn Cuvette Segnænts Reøacgo me CH Reactg:n Rög Cuvette Resiacir" the CH Reacbon Rimg Cuvette Segnents Reøacm me CH Reacbcn Cuvette Reøac•m me CH Reacticn Cuvette Sequents me CH Reacticn Cuvette me CH Reacticn Cuvette SegnæMs the CH Reactü' Rim Cuvette the CH ReactOn Rim Cuvette Segrnents Perform MC物ATELLICA•CH the P«fctm.. Pef«rn.. the Cuvette the Cuætte the Rim Cuvette Rim Cuvette the Cuvnette Rim Cu•ætte Rimg Cuvette Cuvette the Cuvette Cuvette the the performed Capabilities CH 0121 0190 ( 0 1 01 0 80 ( 0 0121 0 90 0 0 01 0 3 00 1121 018 [ 0 the the the activity name A日℃日 activity name Search Text Sea 10 ( narne 1日C「H フ0給20 ) the Resiacir" CH Reacbon Rim Cuvette Segnents Replace the Dilution Ring Cuvette Segments: TEは℃A ( H Schedutod by Operator Scheduled by Operator hedu by 0 Are you sure yau want to exit Diagnostics? CH Cwettes Are you sure you want to enter Diagnostics? As AS 2 20 50 ) the Completing Replacing the Dilution Ring Cuvette Segments Scheduled Autanaticalty Scheduled Scheduled Automatical,9 hed edAまツ 101フ0 84 ( 0 Scheduled by Service Scheduled by Scheduled by Seroce Scheduled by Ser-Oce 0 0 by Seハ0 は℃A H 1121 01308 rte System connected to pcc connected to PCC 1日C CH AS 1121 018 [ 0 1124 01848 Overdue Ckan ロX 1 0 : 055 Rea:tion Bath Rea:tøn Bath Rea:tbn Bath Yes No Completed C001P て2013 4 ( 0 Refreshm the Bath Refreshm WNer Bath As Needed Not P m00 Not pedormed Not Pedormed ATEは℃40H 112 018 As Nee&ed Needed Back Cover With Common ts With Comments With Cot第men ATELLICA-CH 0121 0 90 0 112 020 14 : 10 ATELLICA-CH the Mixer the Mtxet トAs Needed As Needed Enter Enter s Exit Aをy 立0X0く 01.201831823 . Nov 1 05.加182 52 2018 223:03PM 2018 22303PM