Reprocessing PET Quality Check - Biograph mCT and Horizon
This step by step procedure will assist the user in reprocessing PET Quality Check for PET in the event that the incorrect phantom information was entered or if the PET Quality Check fails.
Reprocessing PET Quality Check Biograph mCT and Horizon Description: This step by step procedure will assist the user in reprocessing PET Quality Check for PET in the event that the incorrect phantom information was entered or if the PET Quality Check fails. 1. Pull down Setup à PET Quality Check (Biograph mCT) or Options à PET Quality Check (Biograph Horizon) 2. Dialog box will appear. Select the Configure QC tab. Here you can change any of the phantom information that is incorrect. PET Quality Check Perform and Monitor Configure Phantom Volume 6315 (cc) Isotope Ge-68 Assay Activity 2.44 mCi Assay Time 12:00:00 Assay Date 11-Jan-2011 Cross Calibration 1 Partial Setup Time Alignment Full Setup · Position phantom with laser Position phantom with last saved value QC Schedule day: Mon Tue Wed Thu ID Sat Sun QC Schedule time: 06:30:30 Start View Cancel Help SIEMENS siemens.com/healthineers Healthineers . Reprocessing PET Quality Check – Biograph mCT and Horizon Effective Date: 02/27/2018 | Quality HOOD05162002829044 3. Select the Perform and Monitor tab. 4. Click the Load button to bring up the sinogram list. PET Quality Check Perform and Monitor QC Configure QC -Report Computed ECF 2.8870+007 View Sinogram View Trend View/Print Report Save&Close Post process Start Load Cancel Help 5. Select the QC sinogram you would like to reprocess and click Load. ET Quality Check Perform and Monitor DC Config Saving Quality Check results Quality Check results saved as system defaults QC Results X Sinograms QC Reports PETQCSinogram_30Jul2009_065648.s PETQCReport_30Jul2009_065648 xml PETQCSinogram_29Jul2009_081915.s PETQCReport_29Jul2009_081915 xml PETQCSinogram 29Jul2009 074853.s PETQCReport 29Jul2009 074853 xml PETQC Sinogram 27 Jul2009 082122 s PETQCReport 27 J2009 082122 xml PETQCSinogram_27Jul2009_074314.s PETQCReport 27 Jul2009_074314 xml PETQCSinogram_26Jul2009 083742.s PETQCReport 26Jul2009_083742 xml PETQCSinogram_25Jul2009_102752.s PETQCReport 25Jul2009_102752 xml PETQCSinogram_24Jul2009_074836.s PETQCReport 24Jul2009_074836 xml PETQCSinogram_23Jul2009_165150.s PETQCReport 23 Jul2009_165150 xml PETQCSinogram_23Jul2009 073341s PETQCReport 23Jul2009 073341 xml PETQCSinogram_22Jul2009 072414.s PETQCReport_22Jul2009_072414 xml PETQCSinogram_21Jul2009 075734.$ PETQCReport 21 Jul2009 075734 xml PETQCSinogram_20Jul2009_083755.s PETQCReport 20Jul2009_083755 xml PETQCSinogram_17 Jul2009_084418.s PETQCReport_17 Jul2009_084418 xml PETQCSinogram_16Jul2009 101725.s PETQCReport_16Jul2009_101725 xml PETQCReport_15Jul2009_100637 xml Load Cancel Start Load Cancel Help Siemens Healthineers Services Page 2 of 4 Reprocessing PET Quality Check – Biograph mCT and Horizon Effective Date: 02/27/2018 | Quality HOOD05162002829044 6. Click the Post Process button to reprocess the QC. PET Quality Check Perform and Monitor QC Configure QC Report Computed ECF |2.8870+007 View Sinogram View Trend View/Print Report Save&Close Post process Start Load Cancel Help 7. If the QC passes and all is satisfactory, click the Save and Close button. 8. If you are still having problems with the PET Quality Check, call the Siemens Customer Care Center at 1-800-888-7436. Siemens Healthineers Services Page 3 of 4 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instruc- tions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety infor- mation like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2018
- biograph
- mct
- horizon
- quality check