SMART C Mini CArm Turnover Workbook
This workbook will be used during initial onsite training of the SMART C.
SMART C Mini-C Arm SMART Turnover Workbook Education & Workforce Solutions SIEMENS Your potential is calling. Healthineers* Turnover Workbook | SMART C Mini C-Arm I. Welcome A Message from Nanci Wozniak Welcome to your training experience with Education & Workforce Solutions. I want to personally thank you for choosing Siemens Healthineers. We are honored to have you as our partner and hope you will enjoy this course. Our teams strive to ensure your priorities and needs are met. We sincerely appreciate your hard work around what really matters - the human side of health care. You are our heroes. Thank you for being part of the Siemens Healthineers family and we look forward to our partnership with you. Nani ligh Nanci Wozniak Vice President, Education & Workforce Solutions CX SV USA – Customer Service Organization Siemens Healthineers Clinical Training and Continuing Education Contact Information Customer Care Center - Clinical Application Support: Call 1-800-888-7436 PEPconnect https://pep.siemens-info.com/en-us 2 Turnover Workbook | SMART C Mini C-Arm II. Save Workbook with Notes In order to save this document with your digital notes included, you must use the print to pdf functionality. Clicking the button below will initiate that function. Print to PDF 3 Turnover Workbook | SMART C Mini C-Arm III. Contents I. Welcome ....................................................................................................................................................................2 II. Save Workbook with Notes .................................................................................................................................3 III. Contents ..................................................................................................................................................................4 IV. Course Description ...............................................................................................................................................5 1. SMART C Turnover Workbook........................................................................................................................... 6 4 Turnover Workbook | SMART C Mini C-Arm IV. Course Description Course Objectives Upon completion of this course, you will be able to: Correctly perform start-up and shutdown procedures • Identify the hardware of the Smart C system • Describe the key features of the software • Utilize the Exam Screens to perform patient procedures • Utilize the Image Processing to adjust images • Utilize the system to review and send images to an archive • Describe best practices for patient position • Identify system messaging and cleaning procedures • 5 Turnover Workbook | SMART C Mini C-Arm 1. SMART C Turnover Workbook SMART C Turnover Workbook • System Overview • Hardware • System Software • Exam Workflow • Patient Positioning • System Messaging • System Cleaning/Disinfecting SMART C Notes 6 Turnover Workbook | SMART C Mini C-Arm SMARTC System Startup Status Indicator 6:22 PM 12/17/2018 8:19 AM 1) Turn on the Tablet and C‐arm (~1 Minute) Disconnected Connected • Physical power button on right side of tablet. • C‐arm power button on control panel ID: 2) Admin Verify the C‐arm, Footswitch and Tablet are LOGIN communicating and charged. Login PASSWORD: • Check the status Indicators for connection. • Battery symbols show charge. • No Patient Info= Verify each component is charged. Emergency Exam 3) Login to the Tablet PATIENT ID: X LAST NAME: Enter Username and Password FIRST NAME: • 5 FM BIRTHDAY: 1234 MM/DD/YYYY ACCESSION # : Mike Jordan, M, 2/20/1974 4) Enter patient Information or start emergency exam. x 132 PHYSICIAN : Note: Power off ANATOMY: Mike Trout, M, 2/20/1978 • To search an existing patient enter ID, Last or First Name. Then select the correct exam from the list. performed using 5) Position the Patient onscreen power button. Notes 7 Turnover Workbook | SMART C Mini C-Arm SMART C Power Management X-RAY LOCK -KU --- LUA M |12/17/2018 • Best practice to check battery charge of C‐arm, C 8:19 AM Tablet, and Footswitch prior to starting exam using Control Panel Status Indicator the status indictor. C‐arm Battery Indicator • The C‐arm batteries are swapped during clinical 06/16/2019 POSSO 926 AM use. Two batteries are provided and can be charged in Charger or in Monitor Cart. TEST • The Tablet can be charged using AC Adapter, Docking Station, or Monitor Cart during use. Battery Charge Test Tablet Battery Indicator • Tablet Charge Status is shown on the Tablet Bezel. • Footswitch uses a replaceable AA battery. • Messages are shown to user when C‐arm or Tablet Batteries are low. Tabletop C‐arm Battery Charger Tablet Docking Station and AC Adapter Notes 8 Turnover Workbook | SMART C Mini C-Arm SMART C Sleep and Wake • To conserve power the C‐arm automatically goes to sleep in 5 minutes without user interaction. No SLEEPING PRESS PWR 2 WAKE imaging or button pushes. VCD • To wake the C‐arm press the power button or press the 3 save button on the footswitch. Takes approximately 10 seconds. ON M SLEEP • $ LOW Tablet and C‐arm display messages when the C‐arm is K4 X-RAY LOCK sleeping to provide user system status. DR: 0.00 mGy/min DT: 0.0 min DLC 07/02/2019 • Previously set mode (FLUORO/LOW DOSE/ X‐RAY LOCK) CD: 0.00 mGy DAP: 0.000mGy*cm2 12:34 PM resumes when the C‐arm wakes. • The Tablet can be configured to sleep when using Network System Users Clinical Audit Tools About battery power. Tablet sleep requires the C‐arm to be SETTINGS INPUT turned off or sleeping with no user interaction. Cumulative Dose Limit (mGy) Date format: MM/DD/YYYY 300 • Tablet delay before sleep can be configured to Never or Delay before Data Deletion (hours) Numeric format: 12345.67 System Date Time 5, 15, 30, 60 minutes. Delay before Auto Log Off (minutes) 15 Default View: STANDARD • Wake the Tablet by pushing the external power button. Delay before sleep (minutes) Default rotation: Notes 9 Turnover Workbook | SMART C Mini C-Arm SMART C Tablet • The touchscreen Tablet is used for image visualization and patient data management. • The Tablet is powered by an internal battery charged using the AC adapter, Docking Station, or attaching it to the optional Monitor Cart. Warning: The Tablet contains a fan for active cooling. Do not block M 60kV 350ML the fan exhaust port. Do not place the Tablet inside of a sealed bag. Do not face the fan outlet toward the sterile field or patient. 19 60.350MA Power Audio ports USB Power Button not Fan Connector not for use Port Button for use Exhaust Notes 10 Turnover Workbook | SMART C Mini C-Arm SMART C C‐arm Hinge Release • X‐ray On Trigger Switch Weighs approximately 16 lbs. for portability and easy positioning. C‐arm Head Control Panel • To open or collapse the C‐arm, hold the neck Battery of the C‐arm and press the elastomer hinge Power Button release button on the side of the C‐arm. X‐ray On Button • Ensure the C‐arm Battery latches when it is inserted into the battery pocket. • C‐arm control panel is used acquire images, X‐ray Generator change imaging modes and technique, and perform image manipulations. • Laser aimer used to aid in positioning Image Receptor anatomy for imaging. • Power button uses a long push (~1 second) to prevent accidental power ups. Notes 11 Turnover Workbook | SMART C Mini C-Arm SMART C C‐arm Control Panel • When connected the Footswitch Indicator: Indicator: Button: and Tablet LEDs become Footswitch Connected Tablet Connected Manual kV/mA Button and Indicator: Auto kV/mA illuminated. Button with Indicator: Auto Brightness/Contrast • Yellow LED indicators on the O control panel X‐ray On buttons (4) Indicators: X-ray On flash when imaging. Button: • Mirror Image B LCD display located in the center Button: EG M <-ray On of the control panel shows the mode and system state. Button: Rotate Image Button: • Mode A battery indicator located on the Button: Button: Button with indicator: Button: LCD shows C‐arm battery charge. Contrast/Brightness Save Image Power On/Off aser Crosshair On/Off Notes 12 Turnover Workbook | SMART C Mini C-Arm SMART C Wireless Footswitch Save and Wake • Connects to the C‐arm automatically at boot. • X‐ray On Pedal Button There is no power button on the footswitch. • Wireless footswitch provides X‐ray ON actuation. • Save Button is same as Main Window save. • Press Save/wake button to wake C‐arm from sleep. AA Battery DURACELL • Footswitch only operates when in a horizontal Screw position Battery Cap • Footswitch AA battery changed by unscrewing and replacing battery under cap. Wireless Footswitch Notes 13 Turnover Workbook | SMART C Mini C-Arm SMART C Storage, Setup, and Transport • Hard Shell Case (sold separately) is used to store: 1. C‐arm 2. C‐arm batteries (2), charger, power adapter 3. Tablet 4. Tablet docking station, power adapter 5. Footswitch 6. Head Extension 7. Manual, Quick Cards, and USB • Carry the C‐arm with both hands during setup. • The C‐arm, Tablet, and Footswitch are uniquely paired. Always keep them together. Storage and Transport Conditions Temperature: 0C to 60C (32F to 140F) Relative Humidity: 5% to 95% (non‐condensing) Atmospheric Pressure: 70kPa to 105kPa Ambient Operating Conditions Temperature: 10C to 30C (50F to 86F) Relative Humidity: 25% to 80% (non‐condensing) Atmospheric Pressure: 70kPa to 105kPa Notes 14 Turnover Workbook | SMART C Mini C-Arm SMARTC System Users Rights or Access to: User Type • Admin can change all default settings and view all patient data. View Patient data User Types System Settings View/Edit All patient Edit System Settings System Settings Admin Network Tab in Initiate X-ray Service account can modify all system settings but ~ Config Tab in data • Operator None Only patient View View View Yes data created only view data stored in their account. Operator Reader None All patient • Reader user can view and export all data stored on View View View Nol data the system. Service View Only patient Edit Edit View Yes data created by Service • Operator can view and export data stored within their account. Network System Users Clinica Audi About • System provides access control to features using a USERS Local users: 0 of 20 user authorization system. Admin Admin • All user Passwords can be modified on the system. Service + Service • Service user privileges and higher are required for Operator Operator DIOCM setup. se Reader Reader Notes 15 Turnover Workbook | SMART C Mini C-Arm SMART C Exam Screens • Swiping on the reference window (right side) navigates between user interface screens. SWIPE • Swipe left for Processing or right for Library screen • Swipe twice either direction for Patient Info screen 3 40 kv 130HA SMART C PATIENT ID : 07/03/2019 9:33:20 AM LAST NAME FIRST NAME: PACS USA INC FRAMES (MB) TIME SEX: M RTHD MANDD/WWWY ROSA F CONTRY AUTO (0 MB) 9:33:56 AM V % OFF ACCESSION #: 1-50 (29.1 1B) 93820 AM ANATOMY: PHYSICIAN: IMAGE ANNOTATION - DT LIMIT: 5.0 min T - 1-36 (197 MB) 9:32-46 AM DRAW UNDO UNDO ALL <> -SETTINGS ON OFF OFF LED END EXAM END EXAM M 2 40 kv 130 MA M @ $ 40kv 130 MA M.@ 40KV 130 HA ON M. 40 kv 130 MA D-0.05.may DR:0.00 mGy/min DT:0.1 min CD:0.05 mGy DAP:4.2 mGy*cm* 9:37 AM DR: 0.00 mGy/min DT: 0.1 min CD:0.05 mGy DAP:4.2 mGy"cm* 9:39 AM Patient Info Processing Reference Window Library Notes 16 Turnover Workbook | SMART C Mini C-Arm SMART C Image Processing (In Exam) Edge - IMAGE PROCESSING- C • LIH Edge‐ Automatically applies Edge Enhancement to the last image in future BRIGHT sequences. AUTO V TOFF • Edge‐Applies Edge Enhancement to previously EDGE acquired images shown in the Main Window. -IMAGE ANNOTATION- DT LIMIT : 5.0 min T • RESET Filter‐Allows the user to Enable/Disable K4 recursive filtering. Recursive filtering improves -SETTINGS- the detail of still images but causes motion blur. END OFF OFF OFF QMED METAL LIH EDGE FILTER • X‐Ray Chime button allows the user to modify OFF OFF M 40 kv 139 MA the volume of X‐ray On audible alert. OFF FLUORO SMARTC. EDGE:OFF 07/01/2019 11:37:46 AM DR: 0.00 mGy/min DT:0.2 min FILTER: OFF, METAL: OFF 40 kVp, 130 pA ZOOM:1.0 ROT:0º LIH IMAGE: 36/36 CD: 0.07 mGy DAP: 6.720mGy*cm LIH Edge Filter X‐ray Chime Notes 17 Turnover Workbook | SMART C Mini C-Arm SMART C Exam Review Lock Swap • The current sequence is displayed in Main Window. • The Main Window supports multitouch gestures including two finger rotate, zoom, and pan. • Tap the main window to playback a sequence, adjust brightness, refit, or undo all. • The last image of each sequence is sent as a Last Image Hold (LIH) image that has higher grayscale resolution. • Previous sequences move to the Reference EXAM Window 7/10 / 2019 11:56 :51 AM Y ON • Swap will switch the Main and Reference image $ OFF M 40 kV 130HA OFF FLUORO windows. SMART C 07/10/2019 11:57:18 AM DR: 0.00 mGy/min DT:0.1 min • Lock will retain the image in the Reference 40 kVp, 130 HA LIH IMAGE: 36/36 CD:0.05 mGy DAP:4.2 mGy*cm Window. Notes 18 Turnover Workbook | SMART C Mini C-Arm SMART C Exam Library Selection Save • All sequences taken during an exam are displayed REF 123 steve Foot, Adult (80 kg) Dr. Ivey in the Library by default. 07/01/2019 12:13:22 PM V EO PACS USB • Double tap or drag and drop a library image entry FRAMES (MB) TIME V to display it in the Main Window. 1-36 (19.7 MB ) 1:17:50 PM • Drag and drop the library image entry to the LICH Reference Window 1-58 (35.1 MB) 1:18:00 PM EG LIH • Tap on the Main Window for playback. DRAG 1-30 (17.8 MB) 1:18:11 PM • LIH Images are saved by pressing Main Window 9/30 (4.6 MB) 1:18:11 PM save on the last image of a sequence. END • EXAM Snapshot images are saved by pressing Main Window Save on any other image in the sequence & OFF $OFF M 54kV 214WA OFF FLUORO but LIH. SMART 12345steve EDGE :OFF 07/01/2019 1:18:11 PM DR: 0.00 mGy/min Foot , Adult (80 kg/ LTER: OFF, METAL: OFF DT:9.2 min 07/01/2019 ZOOM: 1.0 ROT :00 48 kVp, 178 pA DAP: 14.638 mmGy*cm 966 1:51 PM • Selection Save used to store checked items. (LIH LIH IMAGE: 30/30 CD:0.16 mGy Images or Sequences) Playback Saved LIH/SEQ/Snapshot Main Window Save Notes 19 Turnover Workbook | SMART C Mini C-Arm SMART C Exam Library Selection Save • 123asteve Foot, Adult (80 kg) All sequences taken during an exam are displayed REF Dr. Ivey 07/01/2019 12:13:22 PM in the Library by default. V EO PACS USB FRAMES (MB) TIME V • Double tap or drag and drop a library image entry to display it in the Main Window. Drag and drop 1 -36 (19.7 MB ) 1:17:50 PM the library image entry to the Reference Window LICH • 1-58 (35.1 MB) 1:18:00 PM Use the slider to manually select an image in a sequence. Press play button to review sequence LIH in real time. DRAG 1-30 (17.8 MB) 1:18:11 PM • LIH Images are saved by pressing Main Window 9/30 (4.6 MB) 1:18:11 PM save on the last image of a sequence. END • EXAM Selection Save used to save checked items including entire sequences. & OFF $OFF M 54kV 214 HA OFF FLUORO • Snapshot images are saved by pressing Main SMART C. Jobs, Steve R: 1. METALLO 07/01/2019 1:18:11 PM DR:0.00 mGy/min Foot , Adult ( 80 kg ) 48 kVP, 178 DT:0.2 min DEMTO BIAL- OFF LIH IMAGE: 30/30 AP 14.638mGy*cm" 07/01/2019 360 1:51 PM Window Save on any image other than the last :0.16 mGy (LIH) during playback. Main Window Save Saved Snapshot Notes 20 Turnover Workbook | SMART C Mini C-Arm SMART C Patient Information PATIENT ID : 12345 X LAST NAME: FIRST NAME: Steve • Patient information is modified by swiping to the Jobs SEX : FM patient information screen. BIRTHDAY: 02/20/1964 • Accession # is used to uniquely identify an exam. ACCESSION #: 31244 • The anatomy field allow users to enter free form ANATOMY: HYSICIAN: text or uses a predefined list of anatomies for Foot, Adult (80 kg) Dr. Ivey ease of use. • The physician field accepts free form text or can be predefined in the settings menu. END EXAM Note: The default anatomies and physicians lists can & OFF SIS OFF M 54kv 21 4HA be modified in the Systems/Settings menu by Admin OFF FLUORO users. SMART 12345eve 07/01/2019 1:18:11 PM DR: 0.00 mGy/min DT:0.2 min 07/01/2019 Foot , Adult ( 80 kg ) FILTER : OFF , METAL : OFF ZOOM:1.0 ROT:00 48 kVP, 178 LA LIH IMAGE: 30/30 OSSO 1:19PM CD:0.16 mGy AP: 14.638 mGy*cm3 Notes 21 Turnover Workbook | SMART C Mini C-Arm SMART C Patient Library Back Button LIBRARY EXAMS: 4 Library v PACS USB Button • PATIENT ID PHYS| PACS DATE - TIME Enter the Patient Library to review exams 07/01/2019 12:13:22 PM saved on the system. obs, Steve Dr. Ivey • Trout, Mike 132 Dr. Basebe 07/01/ 2019 11:37:08 AM Patient Library is entered by pushing the library button after login. Jordan, Mike 1234 Dr. Duffy 07/01/2019 8:26:10 AM • To perform a search for only a specific Jordan , Mike 1234 Frank 06/26/2019 5:04:08 PM patient enter the patient information in the ID, Last or First Name fields. Select the patient from the list and press the Library button. Patient List • EXIT Patient information can be modified in the library. EO |07/01/2019 3:37 PM Notes 22 Turnover Workbook | SMART C Mini C-Arm SMARTC Patient Data Delete and Undelete Data Deletion Time Delete Undelete • If sequences or images are not stored when leaving Network System Users Clinical Audit About 019 10:49:06 AM an exam, they can be recovered using undelete SETTINGS INPUT button. Cumulative Dose Limit (mGy) Date Format: MM/ DD /YYYY FRAMES (MB) TIME 100 Delay before Data Deletion (hours) Numeric Format: 12,345.67 RDSR (0 MB ) 10:51:08 AM • The time the system retains data that is not saved is Set Date/Time Delay before Auto Log Off (minutes) configurable in the settings menu. Never Default View: STANDARD Delay before Sleep (minutes) -58 (35.1 MB) 10:49:29 AM 30 Default Rotation: • When Data Deletion time is set to 0 data is deleted LIBRARY 1-30 (17.8 MB) 10:49:50 AM Reset Anatomies List immediately. Delete All Exams • Reset Anatomies defaults the anatomies list to the END factory defaults. EXAM Y ON M 40kv 130 MA OFF FLUORO • Delete All Exam deletes all patient data on the OFF USED 9 EXAM 01/08/2020 DR: 0.00 mGy/min DT: 0.3 min CD: 0. 10 mGy DAP: 9.1 mGy"cm" system Reset Anatomies Delete All Exams Notes 23 Turnover Workbook | SMART C Mini C-Arm SMART C Image Acquisition (In Exam) Collimate 1) Select Imaging Mode FLUORO / LOW DOSE / X‐RAY LOCK 2) Set Technique Auto kV/µA (default) / Manual kV/µA Reference 3) Collimate as necessary Main Window 4) Position Anatomy and C‐arm Window SWIPE go 5) Acquire an image sequence using one of the 3 Exposure switches on the C‐arm head or using END EXAM 7/1/2019 11:37:38 AM footswitch. Y OFF OFF M FLUORO 40kV 130HA Note: When acquiring a sequence hold X‐ray SMARTC FILTER : OFF , METAL: O 07/01/2019 11:37:48 AM DR: 0.00 mGy/n ZOOM: 1.0 ROT:0. AB kVp, 130 HA CD:0.07 pay DT: 0.8 LIH IMAGE: 36/36 DAP: 6.720 mGy+d switch until the yellow indictor appears on the Tablet and image stabilizes. Imaging Mode Auto kV/µA Manual kV/µA Notes 24 Turnover Workbook | SMART C Mini C-Arm SMART C USB Export, PACS, and Print PACS Export USB Export Main Window Print Selection Print • The system can send sequences and images to PACS once configured. See LIBRARY EXAMS: 2 Jordan , Mike 07/03/2019 10:00:01 AM DICOM setup slide for help. EP PACS USB PATIENT PHYS INFO DATE - TIME FRAMES (MB) TIME • Sequences and images can be Jordan, Mike 123 Dr. James [2:21 07 /03/ 2019 9:33:20 AM exported to USB as .JPG .TIF or .DCM RDSR (0 MB) 10:01:03 AM 1 Page. James 1234 Dr. Hump! [0:0] 07/03/2019 9:31:50 AM • Sequences and images can be sent to HIT 1-50 (29.1 MB) 10:00:30 AM a supported DICOM printer. DRIVE: EXT: .jpg LIN • DRIVE: NOT SELECTED e 1-36 (19.7 MB) 10:00:54 AM Images in the Main Window can be EXPORT: ~ 5 - printed any time by pressing Main EXPORT CANCEL Window Print. END EXAM • Selected sequences and images can ON 9% OFF M 44kV 154 HA be printed from the library screens. FLUORO 07/03/2019 9:59 AM DR: 0.00 mGy/mir DT: 0.1 min DAP: 6.8 mGy+cm2 0 5 - 07/03/2019 CD:0.07 mGy 1 10:01 AM Patient Library Exam Library Notes 25 Turnover Workbook | SMART C Mini C-Arm SMART C Pediatric Patient Positioning and Dose To minimize pediatric and small patient dose • Perform only necessary fluoroscopic examinations. • Image only the anatomical region interest. • Use Low Dose mode which decreases technique to ½ the dose rate of Fluoro mode. • Collimate when possible. • Install Skin Spacer to keep the minimum distance between the patient and X‐ray focal spot. • Use laser aimer for proper positioning. • Use the DT limit, DAP, and RDSR to aid in establishing dose threshold values. Warning: The Smart‐C is for extremity imaging only, • Position the patient extremity as close to the regardless of patient size or age. Whole body detector as possible. pediatric imagining should never be performed. Notes 26 Turnover Workbook | SMART C Mini C-Arm SMARTC Adult Patient Positioning • The Smart‐C is for table top use when used without the support stand. • The thin detector can be slid directly under anatomy. • The system can be quickly rotated around anatomy. • The system is fully operational when it is placed its side or front. • When draped the C‐arm can be placed on a sterile back table when not in use during surgery. Notes 27 Turnover Workbook | SMART C Mini C-Arm SMART C Messaging System • User interface messaging system provides 9:14:18 AM Medium 401, Tablet battery charge is less information on system states and errors. than 25%. Charge tablet. • System Messages are ranked as High, Medium, A or Low severity. High 109, C-arm battery capacity is • Red‐High severity messages inform the user at lower limit. Imaging is disabled. Charge or replace battery. that imaging is unavailable, or shutdown is OK imminent. • EXAM Yellow‐Medium severity messages inform the 7/1/2019 5:38:06 PM DOFF M 40 kv 130 MA user of conditions that, if not addressed, may FLUORO % OFF M FLUORO @ 40kV 130 HLA lead to loss of imaging or shutdown. DR: 0.00 mGy/min DT: 0.0 min C 07/02/2019 DR: 0.00 mGy/min DT:0.0 min D:0.00 mGy DAP: 0.000 mGy*cm2 9:14 AM D:0.01 mGy DAP:0.692 mGy"cm • White‐Low severity messages provide Solid Indicator Messages Popup Message information regarding system status and events. Yellow Medium Priority Red High Priority Notes 28 Turnover Workbook | SMART C Mini C-Arm SMARTC System and Network Logs Network ystem Users Clinical Audit Tools About LOG • Network logging is used for troubleshooting Export Clear log wireless DICOM network. Status is shown while [v3.3.1.0 30.10.2019 21:08:39.354] TIS (Admin): LOGIN [v3.3.1.0 30.10.2019 21:08:39.443] TIS (Admin): Frame server error: An invalid IP address was specified. (521) making connection. • System audit are logs used for troubleshooting hardware and software issues. • Comprehensive system log is for auditing and System Log troubleshooting. ETWORK JETWORK LOG Clear • Login PACS SSID V X • Exam start CARM-Mike O_5G 70% 70% • Data deletion Wireless IP Wireless MAC 192.168.50.31 541379A80F89 • Exam review • All user messages Wired IP Wired MAC • DICOM actions IMAGE STORAGE CAPACITY: 10/30/2019 • Hardware and software issues CAPACITY 9:08 PM G Network Logs Notes 29 Turnover Workbook | SMART C Mini C-Arm SMART C Cleaning and Disinfection Do not spray or allow liquids to drip or run into • Turn off the system and disconnect electronic enclosures. Liquids inside electronic components from the any power source enclosures may lead to electric shock and/or fire before cleaning. hazard. If liquids accidentally spill into the system • Clean any surfaces that have become soiled electronics, DO NOT connect the power cord or battery until the liquids have dried or evaporated. with water or isopropanol. • Use a cotton swab to clean between the blades of the heatsink. Do Not Use: • To clean the Tablet, Docking Station, power Petroleum based cleaning fluids adapters, or other accessories apply Acetone cleaning solution to a cloth and gently wipe. Bleach solutions with over 10% concentration Do not spray liquids directly onto the Tablet, Temperature, chemical, or pressure sterilization electrical components, or connectors. Buffing pads or scrubbing pads Notes 30 On account of certain regional limitations of sales rights and service availability, we cannot guarantee that all products included in this brochure are available through the Siemens sales organization worldwide. Availability and packaging may vary by country and is subject to change without prior notice. Some/All of the features and products described herein may not be available in the United States. The information in this document contains general technical descriptions of specifications and options as well as standard and optional features which do not always have to be present in Local Contact Information individual cases. Siemens Medical Solutions USA, Inc. 40 Liberty Blvd. Siemens reserves the right to modify the Malvern, PA 19355-9998 design, packaging, specifications and USA options described herein without prior Telephone: +1-888-826-9702 notice. Please contact your local siemens.com/healthcare Siemens sales representative for the most current information. Legal Manufacturer Note: Any technical data contained in Turner Imaging Systems this document may vary within defined 1119 S 1680 W. tolerances. Original images always lose a certain amount of detail when Orem, UT 84058 reproduced. Phone: 866-870-2022 turnerxray.com Siemens Healthcare Headquarters Siemens Healthcare GmbH Healthcare Sector Henkestr. 127 91052 Erlangen Germany Phone: +49 9131 84-0 siemens.com/healthcare Published by Siemens Healthcare GmbH · HOOD05162003123771 · Printed in USA · ©Siemens Healthcare GmbH, 2020
- SMART C