Smart_C DICOM Connectivity Setup

This job aid will guide the user through the correct steps for setting up the
DICOM connectivity between the hospital and the Smart-C

Smart-C DICOM Connectivity Set Up SMART-C The following guide can be used to to install, configure, and test DICOM network connections on the Smart-C in a typical healthcare environment. The Smart-C supports the following DICOM features: DICOM Store: Transmits DICOM patient and image data to PACS servers for long term storage and archival. DICOM Modality Worklist: Allows users to import patient data from the network prior to an exam to reduce text entry errors and improve workflow. The system may be configured to allow DICOM data to be sent through a wireless network when in range, and/or through a wired network connection when the tablet is placed in the tabletop docking station. 1) Prerequisites: Confirm the customer DICOM network supports the following: a. Wired network requirements i. Ethernet Network supporting at least 10Mbit/s or 100Mbit/s wired connectivity ii. Internet Protocol version 4 (IPv4) iii. Dynamic Host Configuration Protocol (DHCP) (1 IP address) b. Wireless network requirements i. Wireless communication protocols: 802.11ac, b, g, n ii. Wireless Authentication: WPA2-PSK iii. Internet Protocol version 4 (IPv4) iv. Dynamic Host Configuration Protocol (DHCP) (1 IP address) 2) Complete the required values below for DICOM installation. Obtain the following information from the IT department or systems administrator and complete the form below. Note: some data is optional and not required as indicated. SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Job Aid | DICOM Connectivity Setup Effective Date: 11/13/2020 | HOOD05162003145888 Setting Type Value Description Required for: Mo- Local AET dality Worklist Displays the DICOM Application En- DICOM Store tity Title for the local server. Station Optional Displays the unique name used Name Value De- to identify the system in the list fault: Smart- of C available destination devices. Gateway Optional Value Enter the Gateway IP address. This Default: 0.0.0.0 value is optional and used in large installa- tions. Required for: Store AET DICOM Displays the DICOM Application En- Store tity Title for the STORE server. Default: 0.0.0.0 Required for: Store IP DICOM Enter the unique IPV4 address as- Store signed to the configured STORE Default: 0.0.0.0 server. Required for: Enter the port number assigned to Store Port DICOM the server running on the Tablet. Store The default value is 104. Valid val- Default: 104 ues are 1- 65535 Required for: Mo- Displays the DICOM Application En- MWL AET dality Worklist tity Title for the Modality Worklist Default: Blank (MWL) server. Required for: Mo- Enter the unique IPV4 address as- MWL IP dality Worklist signed to the configured Modality Default: 0.0.0.0 Worklist server. Required for: Mo- Enter the port number assigned to MWL Port dality Worklist the server running on the Tablet. Default: Blank Valid values are 1-65535 SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Page 2 of 6 Job Aid | DICOM Connectivity Setup Effective Date: 11/13/2020 | HOOD05162003145888 3) Setting Up Network Connections: a. Connecting to a wired network i. Turn off the C-arm. ii. Login to the Tablet as User = Admin and password = admin or User = Service and password = service. iii. Go to the Settings/Network screen. (Figure 1) iv. Connect the power adapter to the back of the docking station. v. Connect the ethernet cable from the DICOM network to the docking station. vi. Place the tablet on the docking station and confirm that a wired IP was assigned to the tablet form the DICOM network DHCP server. vii. Confirm the Ethernet box is green and shows the IP address assigned, MAC address, the wired connection status bar. b. Connecting to a wireless network i. Turn off the C-arm. ii. Login to the Tablet as the Admin or Service User. iii. Go to the Settings/Network screen (see figure 1) iv. Press on the white Network box that says “PACS SSID” this will open a list box showing the wireless networks detected. Select the wireless DICOM network and press the check mark. v. To connect to Hidden network select Hidden from the list box. Enter SSID. vi. Enter the wireless network password in the same box. vii. Remove the Tablet from the docking station if connected. viii. Confirm the wireless box shows the wireless IP address assigned, MAC address, and showing connected with a green background on the wireless connection status bar. When connected to the wireless DICOM or C-arm network the signal strength is shown on the box. See Figure 2. ix. Confirm that the Network Log for the wireless network shows “Connection Complete”. Note: To delete the wireless settings for a network including password push the back arrow when the network is selected. The check when pushed will force the reconnection to the wireless network and update the password after deletion. SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Page 3 of 6 Job Aid | DICOM Connectivity Setup Effective Date: 11/13/2020 | HOOD05162003145888 Network System Users Clinical Audit Tools About DICOM Local Local AET Station Name STORESCP DICOM Store Server Store AET Store IP Store Port ORTHANC 127.0.0.1 65000 DCM Verify DICOM Modality Worklist Server MWL AET MWL IP MWL Port ORTHANC 127.0.0.1 4242 DCM Verify NETWORK NETWORK LOG Clear PACS SSID CARM-Mike O_5G 70% Wireless IP Wireless MAC 192.168.50.31 541379A80F89 Wired IP Wired MAC IMAGE STORAGE CAPACITY: |10/30/2019 USED NO EXAMS CAPACITY 9:08 PM C Figure 1 4) Configuring Modality Worklist and DICOM Store. (Figure 2) a. Enter the values obtained in DICOM setup information table above on the Network screen. b. The Local AET, server IP address, server AET, and server port are required for each connection type. c. The system uses a dynamic firewall to open only during transmission/ receipt of patient data. The tablet does support response to ping requests when connected to the external DICOM network. d. Confirm DICOM Store server connection by pressing the “DCM Verify” button on that row. The button will turn green when it completes the DICOM verification successfully. e. Confirm DICOM Modality Worklist server connection by pressing the “DCM Verify” button on that row. The button will turn green when it completes successfully. f. After successfully confirming the Modality Worklist DCM Verify exit the Network settings screen. SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Page 4 of 6 Job Aid | DICOM Connectivity Setup Effective Date: 11/13/2020 | HOOD05162003145888 Network System Jsers Clinical Audit Tools About DICOM Local Local AET Station Name Gateway STORESCP Smart C 0.0.0. DICOM Store Server Store AET Store IP Store Port ORTHANC 10.0.0.10 4242 DCM Verify DCM Verify DICOM Modality Worklist Server Green=Success MWL AET MWL IP MWL Port Red=Failed ORTHANC 10.0.0.1 4242 DCM Verify NETWORK NETWORK LOG Clear PACS SSID [14:23:11; Turner] Connection Complete Wireless Network Turner [14:23:11; CARM-MikeO] Disconnected [14:23:11; CARM-MikeO] Disconnecting Green=Success 5Turner [14:22:01; Turner] Connecting 99% [ 14:05:40; CARM-MikeO] Connection Complete Red=Failed [14:05:30; CARM-MikeC] Connecting Signal Strength Wireless IP Wireless MAC 4:53.13; Ethernet] Connected 10.0.0.227 F834417EBD23 (11:53:01; Ethernet] Disconnected [11:52:27; Ethernet] Connected E Ethernet [11:52:20; Ethernet] Disconnected Wired Network Green=Success Wired IP Wired MAC 10.0.0.244 448E12037DAC Red=Failed IMAGE STORAGE CAPACITY USED NO EXAMS LE| | |03/10/2020 2:35 PM G Figure 2 5) Confirm a list of patients is shown on the Modality Worklist screen. 6) Confirm the successful transmission of the DICOM images and patient data entered to the PACS system. SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Page 5 of 6 Job Aid | DICOM Connectivity Setup Effective Date: 11/13/2020 | HOOD05162003145888 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter co llectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and c autions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be comme rcially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com SIEMENS Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Job Aid Effective Date: 05/XX/2020 | HOOD0516200308 Healthineers Page 6 of 6

  • SMART C