Somaris X Shutdown Procedure

This document will outline the properly recommended shutdown protocol for SOMATOM go. & SOMATOM X.cite systems.

SOMATOM go & X.cite Shutdown Procedure Job Aid SOMATOM go & SOMATOM X.cite.; Somaris X This document will outline the properly recommended shutdown protocol for SOMATOM go. & SOMATOM X.cite systems. • Shut down daily [full system off] ** • Restart/reboot when prompted ** • Checkup/Calibration at 75 minutes – prompted ** • DO NOT CLICK “DON’T SHOW THIS AGAIN’ • Calibration recommended if system has been idle for an extended time period (2 hours or more) and/or If artifacts present or change in temperature ** Please see instructions listed below in this document. Note: Every system is different. Please follow instructions from the Clinical Service Engineer (CSE) and or your Service Customer Care Team. System Startup Successful startup means that the system connects power to all system components such as comput- ers, the table, and the gantry. Full functionality is guaranteed when the Home screen is displayed and the keys on the control box light up. The On/Off button in the gantry Connector Box located in the right gantry stand will turn on the sys- tem. Note: If a wall switch is installed, there is an On/Off button to power the scanner. The line connection box (LCB) power switch is also an option. Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 03/22/2021 | HOOD05162003170812 Healthineers SOMATOM go. & X.cite Shutdown Procedure Effective Date: 03/22/2021 | HOOD05162003170812 System Check up To ensure safe operation, best image quality, and to correct any system irregularities, it is recommended to perform a Check up procedure in regular intervals. If a Check up is not executed after 24 hours a pop-up message will appear, but can be cleared. If the Check up is still not performed at 36 hours the message cannot be cleared until the Check up is initi- ated. Steps to perform a Check up: 1. Select System Check from the Home screen or Check up in the Settings menu. • The Check up window opens. Check up, System Check 2. Select GO. • You are prompted to press Start. For optimum image quality and system safety, please run the Checkup procedure. Exit the scan room. The table will move automatically. X-rays will be produced during the Checkup procedure. Repeat GO cancel Checkup has started. Press the GO button to continue. 3. Press Start on the control box. • Scanning is initiated. The system Check up begins. + SIEMENS Healthineers rs ...... + Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 03/21/2021 | HOOD05162003170812 Healthineers Page 2 of 4 SOMATOM go. & X.cite Shutdown Procedure Effective Date: 03/22/2021 | HOOD05162003170812 System Shutdown Shutting down the system will close all open workflows and save the results to the Short Term Storage (STS). For efficiency, you can exit all running examinations and close all applications before proceeding. You can shut down the system from the Home screen: 1. On the access bar, click the Home icon. Q Results (fraud . Johnson, Elise ------ Joyner, Felicia . Head, Marcelo . Ford, Manuel · Chaney, Aurelio ---- · Mayo, Pauline Instances . Jackson, Denver Note: The computer terminal should be restarted every 24 hours via Home screen, then clicking Restart System to assure best system performance. The entire system should undergo a complete controlled shutdown once every 7 days. 2. On the Home screen, click the arrow head below the Shut down icon. O 3. Select Shut down System. Note: If a workflow is open or a job is running that uses local resources, a warning message is displayed. To shut down the system, click Confirm. Select Yes, and a dialogue box shows the shutdown progress. Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 03/21/2021 | HOOD05162003170812 Healthineers Page 3 of 4 SOMATOM go. & X.cite Shutdown Procedure Effective Date: 03/22/2021 | HOOD05162003170812 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Head- quarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 03/21/2021 | HOOD05162003170812 Healthineers Page 4 of 4