PEPconnect

SOMATOM go.All, go.Top Job Aid How to use CARE kV grouping

job aid that outlines steps for VA30 software on SOMATOM go.All, go.Top for CARE kV grouping

How to Use CARE kV grouping SOMATOM go.All/SOMATOM go.Top Adapting the kV value for each patient offers significant dose reduction potential. However, changing the kV impacts the Hounsfield Unit value in the resulting images. In CARE kV, you can add scan entries to a Group to set the same kV value for all scan ranges at once, for example, to perform a three-phase liver examination. This value is applied to the entire scan range and does not modulate during the scan. In 5 easy steps, this document guides you through the workflow of how to turn on CARE kV grouping for SOMATOM go.All and SOMATOM go.Top. Please note: This Job Aid reflects software version VA30. • CARE Dose4D and CARE kV are combined where the CARE kV license is available. • The CARE Dose4D & CARE kV is set to Full. • All scan ranges that are added to the CARE kV group are scanned with the same kV value while the other settings are retained. • For scans with a CARE kV group with different limits, CARE kV selects the kV values within the defined limits for all scans of this CARE kV group. • Only one group can be created in one study. 1. How to Turn on CARE kV grouping: • On the timeline, right mouse click (or double click) on the scan range and choose Show Parameters to open. Show Parameters Delete Repeat Open Edit Mode Create Image with Reference Lines 4.95 16 2. Open the Dose tab. Scan Favorites& General Scan Dose iming Contig Physio Scan Contrast Scan/Rein CARE Dose4D & CARE KV KV Eff. mAs CTDIvol [mGy] DLP I'mGy *cm] Shaped Filter Dose Notification ref. CTDI [mGy] Dose Notification ref. DLP [mGy*cm] Non Contrast Liver Off 120 129 Standard 0.00 0.0 Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 07/21/2020 | HOOD05162003105672 Healthineers Job Aid | CARE kV grouping Effective Date: 07/21/2020 | HOOD05162003105672 3. On the CARE Dose4D & CARE kV tab, go to the drop down under, change to Full. Scan Favorites General Scan Dose Timing Scan/Recon CARE Dose4D & CARE KV KV Eff. mAs CTDIvol [mGy] DLP [mGy*cm] Non Contrast Liver Off 120 129 (32 cm) 10.84 CARE Dose4D & CARE KV Note: If CARE Dose 4D & CARE kV are set to “Manual” or “Off”, CARE kV Full - Grouping will not be available. Manual kV Full 4. CARE kV grouping now becomes available. Turn “On” grouping if not already set. CH Scan Favorites General Scan Dose Timing Config - Scan/Recon CARE Dose4D & CARE KV CARE kV Grouping CARE KV CARE KV Quality optimized for ref. mAs @120 kV KV Non Contrast Liver Full On Non-Contrast 129 120 5. Verify that the grouping is turned “ON” for all your scans. This can be done in three ways. • Either by checking the scan parameters in Timeline scan parameters: Scan Favorites General Scan Dose Timing - Scan/Recon CARE Dose4D & CARE KV CARE kV Grouping CARE KV CARE KV Quality optimized for ref. mAs @120 kV kV Non Contrast Liver Full On Non-Contrast 129 120 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 07/21/2020 | HOOD05162003105672 SIEMENS Healthineers Job Aid | CARE kV grouping Effective Date: 07/21/2020 | HOOD05162003105672 • Or in the Protocol Selection scan parameters. Abdomen Multi Phase [factory] Protocol Description -Indications: masses, tumor, lymphoma, lymph nodes of the liver, pancreas, gall bladder, spleen, kidneys, interstine, w and wo CM, 3 phases preconfigured Gran madlac: DantinaSpiralAdultAhdamon CH Scan Favorites General Scan Dose Timing Config Recon Favorites General Recon Recon&GO Image Impression Recon Box Scan/Recon CH Dose Scan/Recon CARE Dose4D & CARE KV CARE KV CARE KV optimized for CARE KV Quality Grouping ref. mAs @120 kV KV mA Topogram Sn 100 55 Non Contrast Liver Full On Non-Contrast 129 120 Arterial Phase Abdomen Full On Soft Tissue with Contrast 129 120 Venous Phase Abdomen Full On Soft Tissue with Contrast 129 120 • Or in Exam Designer scan parameters. Exam Designer Scan Protocols 1. Select Protocols 2. Modify Protocols 3. Confirm Changes CH Protocol Scan Favorites General Scan Dose Timing Config Recon Favorites General Recon Recon&GO Image Impression Recon Box Physio Recon Find/Replace Scan/Recon Dose Scan/Recon CARE Dose4D & CARE KV CARE KV CARE KV Quality Grouping CARE KV optimized for ref. mAs @120 kV kV Quality ref. mAs mA Abdomen Multi Phase [f. Topogram Sn 100 55 Non Contrast Liver Full On Non-Contrast 129 120 129 Arterial Phase Abdomen Full On Soft Tissue with Contrast 129 120 129 Venous Phase Abdome Full On Soft Tissue with Contrast 129 120 129 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 07/21/2020 | HOOD05162003105672 SIEMENS Healthineers Job Aid | CARE kV grouping Effective Date: 07/21/2020 | HOOD05162003105672 Note: When CARE Dose4D & CARE kV is changed to Manual kV, you will get a Dose Modulation window pop-up. CARE kV optimized for is still available to optimize your contrast. Dose Modulation x ? Dose Modulation is switched on. CARE kV Quality ref. mAs @120 kV is restored for the reference patient size and organ characteristic: Abdomen. OK Scan Favorites General Scan Dose Timing Scan/Recon CARE Dose4D & CARE KV CARE KV CARE KV Quality optimized for ef. mAs @120 kV KV Non Contrast Liver Manual kV Non-Contrast 129 120 When it is set to Off, you will get a Dose Modulation window pop-up stating that the modulation has been turned off and it will direct you to set your own technique for the appropriate patient size and body part. Dose Modulation x ? Dose Modulation is switched off. Please adjust the mAs value according to the individual patient size. OK Ci Scan Favorites & | General Scan Dose Scan/Recon CARE Dose4D & CARE KV kV Eff. mAs Non Contrast Liver Off 120 129 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 07/21/2020 | HOOD05162003105672 SIEMENS Healthineers Job Aid | CARE kV grouping Effective Date: 07/21/2020 | HOOD05162003105672 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 07/21/2020 | HOOD05162003105672 SIEMENS Healthineers

  • SOMATOM go.All
  • SOMATOM go.Top
  • CARE kV grouping
  • exam split