SOMATOM go. platform Printing Job Aid
This job aid provides instructions on how to print patient scan images when operating a SOMATOM go. platform scanner.
Disclaimer:
Please note that the learning material is for training purposes only!
For the proper use of the software or hardware, please always use the Operator Manual issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. Functionalities described in the material or parts of this functionality may not yet be released for customers and not yet be commercially available in every country. Due to regulatory requirements, the future availability of said functionalities or parts thereof in any specific country is not guaranteed. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients and institutions that are used in these instructions for use are entirely fictional. Any resemblance to names of existing people or organizations past or present is entirely coincidental. All parameters and images shown in this online training are examples. Only the parameters displayed by your system are definite. Configuration-dependent designations, such as names of drives, network nodes, and databases, that are used in these instructions are usually not the same as the designations to be found on a particular installation of the system in a clinical environment. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
SOMATOM go. platform Printing Job Aid This document will cover Print Configuration options, Print Tools, and three ways to print patient scan images. The three ways to print images are via: • Auto Tasking in the User Interface • Local Data • “Save and Print All” in CT-Dental, CT-Calcium Scoring, and MM RT-ImageSuite Print Configuration Options 1. From the Print tab, you can change the Print Layout by selecting the Print Layout Gallery icon. SIEMENS meduser ? = Scan 2 = Recon 1 Ci Shaffer Vonda * 11/01/1967 (50) F e588ccf9-e941-4681-b9f CT View&GO Thumbnails CT View&GO Viewing Add Study 3D Current Print 1 CT, 20/04/2017 Pulmonary CTA(Adult) Topogram 070 7.20 omoniusing 10 00 Pulmonary CTA 5 00 Close Favorite Tools BVSECTA 500 Bv36 83 sag (0) Ditanos ROI Circle Parallel Shipthol Export Stiow VI Line Ranges Image 30 Bred 83 ax (11) 09:30 HOOD05162002589413 Effective Date: 05/26/2017 Page 1 SIEMENS siemens.com/healthineers Healthineers SOMATOM go. platform Printing Job Aid Print Configuration Options 2. To open the Printing Properties, select the Configuration icon at the top right of the Print tab. 3. Confirm that the Default Basic and Advanced settings fit your needs. 4. Select Save as Default Settings then OK to apply changes to current and future prints or Apply then OK to apply changes to the current job. SIEMENS meduser = Scan 2 = Recon 1 Shaffer Vonda * 11/01/1967 (50) F e588ccf9-e941-4681-bet. CT View&GO Thumbnails CT View&GO Viewing cocococo 3D Current Print 2 Printing Properties CT, 20/04/2017 Configuration Pulmonary CTA(Adult) pan Drop Attow Printer . Microsoft Print to F Y Layout 4x6 Printer . Microsoft Print to F Paper Size Letter Copies Paper Size Letter Basic Advanced 3 Topogram 0.70 Tr20 Premonitoring 10 00 Copies Auto-Repack Smart Image Text Collate Sheets Full text on first image Number of Print-outs: 0 Remove Printed Sheets Customized text on first image Monitoring 10 00 Auto-Save Printed Sheets Pulmonary CHA 5,00 V · Portrait Orientation Automatic Image Text Size Favorite Tools Landscape Orientation Fixed Image Text Size 10 pt BY36 53 sag 191 Distance ROI Circle Parallel Snapshot Export Show itt Line Ranges Image 30 Header Footer Save as Default Settings Restore Default Settings Br40 83 ax [13] 4 OK Apply Cancel HOOD05162002589413 Effective Date: 05/26/2017 Page 2 SIEMENS siemens.com/healthineers Healthineers SOMATOM go. platform Printing Job Aid Print Tools Zoom Pan allows you to zoom in and out and pan images in the Film Sheet Drag & Drop allows you to drag and drop series into the Film Sheet Text and Arrow use these options to add text or callout specific details with arrows directly on the Film Sheet Printer allows the selection of connected printers Paper Size allows the selection of different film size options SIEMENS meduser » ? ¢ = Scan 2 = Recon 1 ci Shaffer, Vonda * 11/01/1967 (50) F e588ccf9-0941-4681-b9f CT View&GO Thumbnails CT View&GO 52101 Viewing 3D Current Print . C.T. 20/04/2017 T "Pulmonary CTA(Adult) Text AlTOW Printer . Microsoft Print to F Paper Size Letter Premorstoring 10 Copies Print Number of Print-outs: 0 Monitoring 10 06 Bv36 53 x [5] V Favorite Tools Bv36 Ss cor in Datance ROI Cède Parallel Snapstict Show in Line Ranges Image Print T Zoom Drag & Text Arrow Pan Drop Printer Microsoft Print to F Paper Size Letter Copies Print Save mage Number of Print-outs: 0 HOOD05162002589413 Effective Date: 05/26/2017 Page 3 SIEMENS siemens.com/healthineers Healthineers SOMATOM go. platform Printing Job Aid Print Tools Right-click on the center Film Sheet to open the drop-down menu with tools for the displayed Film Sheet. SIEMENS meduser Text = Scan 2 = Recon 1 ci Shaffer, Vonda * 11/01/1967 (50) F 6588cc/9-e941-4681-591. CT View&GO Thumbnails CT View&GO Arrow Viewing 3D Text Current Select Series Print 71 AmTow Select Senes CT, 20/04/2017 T All images Succeeding be Pulmonary CTA(Adult) Reverse Series Reverse Series Zoom Drag & Tent Pan Drop Cut (Chf+X) Printer Microsoft Print to Copy (CHI.C) Cut (Ctrl+X) Paper Size Letter Paste Copies 1% Delele (Dal) Image Text Copy (Ctrl+C) Zoom/Pan Number of Print-outs: 0 Reset Window Paste (Ctrl+V) V 11 Delete (Del) Favorite Tools Image Text Distance Hol Cimme Parasol Line Ranges Zoom/Pan Reset Window Save Image(s) You can also right-click on a preview Film Sheet to open a drop-down menu with tools. SIEMENS meduser = Scan 2 = Recon 1 CF : 1/01/1967 (50) F 6588ccf9-0941-4681-bit. CT View&GO Thumbnails CT View&GO Viewing 3D Current Delete (Del) Print CT. 20/04/2017 T Print Selection Delete (Del) Printer . Microsoft Print to F Print Selection Paper Size Letter Repack Repack Sopegram 070 7120 New Sheet Select All (Ctrl+A) New Sheet Number of Print-outs 0 Save Image(s) Select All (Ctrl+A) V Favorite Tools Save Image(s) E com HOOD05162002589413 Effective Date: 05/26/2017 Page 4 SIEMENS siemens.com/healthineers Healthineers SOMATOM go. platform Printing Job Aid Printing from the User Interface via Auto Tasking 1. Auto Printing options can be modified within the Recon Parameters of the Auto Tasking tab. 2. Auto Printing can be turned on or off and the layout of the print can be determined. SIEMENS LE Scan 2 E Recon 1 Shaffer, Vonda Cf - 1/01/1967 (50) F 9580cc19-0041-460 1-bor Pulmonary CTA (Adult) VAIDA RELEASE LA16 78466_12412 SAIBA FELEASE VAJOL 6E1 20170484.19493 Recon Ranges Topogram Premondoning Premonitoring TOP 1 Precondinon Monitoring Monkobra (32 cn) 20 01 20 0 110 Pulmonary CTA ROI Definition . Pulmonary CTA Reconstruction A Close 110 De TRATTA 1 Rocon Favontes Gonoral Rocon Image Impression Recon Box Physio Anto Tasking Rocon&GO Scan Recor Auflo Recon Recon Ref Requested Body Part Transfer Auto Lines Nodes Auto Auto Pinta Printing Layout RAmonary CIA 5.00. Pulmonary CIA 5.00. CHEST 2 30 Activate Monitoring scan GO 09:23 3. The configured series will be sent automatically to CT View&GO and load into the Print step. Select the Print tab. SIEMENS meduser › ? # # = Scan 2 = Recon 1 Shaffer, Vonda . 11/01/1967 (50) F 6588ccf9-6941-4681-b9f. CT View&GO CT View&GO Thumbnail Viewing Add Study Current 3 CT, 20/04/2017 Pulmonary CTA(Adult) 3D han drag data from Seites Navigator Print Distribution V 11 Topogram 0.70 1120 Monitoring 10.00 Purmorility CTA 5.00 Favorite Tools Distance ROI Circle Parallel Snapshot Export Show. in Line Rangne Image 30 drog data from Ser les Navigator. Please di og data from Series Navigator, Presse diog data from Series navigator. 161 Des ES 9CAB 2009 5 CTA, 5 00 09:27 HOOD05162002589413 Effective Date: 05/26/2017 Page 5 SIEMENS siemens.com/healthineers Healthineers SOMATOM go. platform Printing Job Aid 4. Select Print. SIEMENS administrator ? @ @ Scan 5 = Recon 0 O Burks, Kelly * 09/11/1978 F 81di6dc0-0541-4873-8636-24 CT View&GO Thumbnail Viewing 3D Current Print CT. 20/04/2017 Thorax CTA(Adult) T Printer . Microsoft Print to F Paper Size Letter 4 Print Num outs 0 x Favorite Tools BAR CTA 5:00 Unage 11:53 CT View&GO – Manual Printing 1. Right-click on a series from the Series Navigator that you would like to print. 2. Select Print Series. The series images will auto-populate the Film Sheet and preview Film Sheet area. 3. Select Print. All series listed in the preview Film Sheet area will print. SIEMENS administrator - = Scan 5 = Recon 0 Burks, Kelly - 09/11/1978 F 81df6dc0-0541-4873-8836-24 CT View&GO Thumbnails Co-COCOCE 6 Viewing 3D Current Print CT. 20/04/2017 Thorax CTA(Adult) Drag & Drco Printer Microsoft Print to F Paper Size Letter 3 Imnapa Number of Print-outs 0 x 1 Favorite Tools Export Series Print Series 2 Barges unage Delete Renome 11:53 Note: If “Remove Printed Sheets” is selected within the Basic Default Settings, after selecting Print the entire print job is sent to the printer and the Film Sheet displays are empty. Refer to Print Configuration options on page 2. HOOD05162002589413 Effective Date: 05/26/2017 Page 6 SIEMENS siemens.com/healthineers Healthineers SOMATOM go. platform Printing Job Aid Printing from Local Data 1. From the Local Data, select the series you would like to print. 2. Select the Print icon located in the Local Data toolbar. A print screen opens. SIEMENS Scan 5 = Recon O o Patient Browser Scheduler Local Data Q More Fithors 2 Mapor Capas Canec Results (A results, 11 55 03) Series Pabwordt Shaly Dalu Saxly Descrit Accession Nu Workflow Burks, Kelly 209/11/1978. F 1 20/04/2017 1 Thorax CTALA Shaffer, Vonda "11/01/1907.F Russo, Evan +24/07/1930. M Dual4ct:2-4784 4cuad 9753-611000082dc3 Foley, Ronnie *11/01/1967.M Instances 30006031-0504-4907-360e 75d1330ifb) . Merrill, Kelli *07/10/1982, F 6811/208-9515-40a2-9061-701096630080 Rose, Gus -3007/1995. M 1a8a7oef Jc10:4487:453c 20ef0oned653 Forbes, Mickey *2505/1930. M The selected dala opened an wund only 11:55 3. Once you have confirmed all settings, select Print. SIEMENS administrator 1 E Scans = Recon O O. Burks, Kely - 09/11/1978 F 81d16dc0-0541-4873-803b-24. Thumbnails Printing Print Add Study C Current CT, 20/04/2017 Thorax CTA(Adult) Printer . Microsoft Print to F Paper Size Letter 3 T Tools Donch Alion 11:55 Note: Print Configuration options and Print Tools are modified using the steps outlined at the beginning of this document. HOOD05162002589413 Effective Date: 05/26/2017 Page 7 SIEMENS siemens.com/healthineers Healthineers SOMATOM go. platform Printing Job Aid Printing from Special Applications – CT-Dental, CT-Calcium Scoring, or MM RT-ImageSuite 1. After an Evaluation has been performed (Dental in this example), select Save and Print All. A print screen opens. SIEMENS = Scan 5 # Recon 0 O Merrill, Koth administrator -07/10/1982 F 68111200-9515-4012-9061-7. CT Dental Thumbnails CT Dental SOMATOM go Up 30 SOMATOM go Up 26 Current 1 CT. 07/04/2017 Dental(Adult) Preset Dental Preset aut x o Findings Navigator Vahie Source MIO Fuera inages Tools seral trage Orientatos -- Merril, Keil HRA Graphics Meni, Kel 201122de-5515 40-2-9001-7.1694 Bragahet 3000 11:58 2. Select the Print tab below the application tab. 3. Select Print. SIEMENS administrator E Scans = Recon o O Memil, Kell -07/10/1982 F 68111200-9515-4012-9061-7. Thumbnails CT Dental Q + CT Dental Print 2 Current . ST. 07/04/2017 Drag & Drop Printer B. Microsoft Print to i v Paper Size Letter 3 Findings Navigator Nome Value Source Tools Rietel inieshet Print Expen 11:59 Note: Print Configuration options and Print Tools are modified using the steps outlined at the beginning of this document. HOOD05162002589413 Effective Date: 05/26/2017 Page 8 SIEMENS siemens.com/healthineers Healthineers
- somatom go.
- somatom go
- somatom
- somotom
- somatom go. up
- somatom go.now