SPECT Dose Report Job Aid
SPECT Dose Reports are created for all acquisitions each time radiopharmaceutical information is entered in an activity. This feature is available only for the MI Apps VB22 software version and up.
SPECT Dose Report Job Aid SPECT & SPECT/CT SPECT Dose Reports are created for all acquisitions each time radiopharmaceutical information is entered in an activity. To create a SPECT Dose Report: 1. Enter the Isotope, Pharmaceutical, Dose and Injection Date and Time in the Series Information tab of the Acquisition Activity. The isotope and pharmaceutical combination must be in the table at the end of this document in order for the SPECT dose report to be generated. bxhchgg. 19.02.19-16:28:24-STD-1.3.12.2.1107.5.6.1.4702, 2/19/2019 File Tomo Acquisition View Template Workflow Activities Help AIR Acquisition Analyzer Display/Analysis Data Selector Patient bxhchgg Acquisition Display Detector 1: Detector 2: Study. 99m Technetium 99m Technetium Series: Tomo Acquisition Summary Series Information| Camera Parameters |Stop Conditions Series: Flexible Display Technologist Reading Physician: Hard Copy Referring Physician: kCounts kicounts Organ: Radiopharmaceutical [ Enable Quantitative cquisition sotope: Pharmaceutical: Radiopharmaceutical Isotope: Pharmaceutical: 99m Technetium 99m Technetium Injection Date and Time: mCI Injection Date and Time: Dose: 0 2/19/2019 12:00:00 AM Can Dose: 0 mCi 2/19/2019 12:00:00 AM Interventional Drug Interventional Drug: Injection Time (hh:mm:ss): 0 0 : 0 nject Supine Dose (mg): 0 64 X 64 1.0 Complete Suspen Tc99m-NMG Both Detectors Setup Prepare Acquisition Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 ) SIEMENS Master Template Effective Date: 3/11/2019 | HOOD05162002998095 Healthineers Job Aid Effective Date: 17/01/2020 | HOOD05162003042654 Job Aid | SPECT Dose Report Job Aid Effective Date: 17/01/2020 | HOOD05162003042654 2. To automatically transfer a dose report to DICOM nodes, click the Setup icon in the lower left corner of the workflow. It will display in the Completion Setup window. 3. From the Transfer Setup area, select the Transfer Dose Report Series checkbox. Note: Even with the dose report checkbox selected, the dose report will not be transferred without previously entering the radiopharmaceutical information. A dose report can also be transferred through the Patient Browser. If a dose report is transferred from a syngo MI Workplace to one with an older software version, the transfer will fail. Completion Setup x Series Save Schedule Background Printing upon Completion The list below contains all the series data that is created by the workflow. All series that are checked will be saved in the database upon completion of the workflow. Schedule Transfer to DICOM Node upon Completion · Save PET as NM Transfer Setup Save PET as PET Upon Completing the Workflow, Transfer the Study/Studies Transfer Dose Report Series Convert VOIs to RT . Transfer the Series Transfer IQ SPECT DC Series Data Selector Dynamic Acquisition To DICOM Nodes: BAY1 Acquisition Output Series(SERIES/NM/DYNAMIC) BAY2 Flexible Display BAY3 Hard Copy BAY4 Save Screen BAY6 Schedule Archive to Device upon Completion Archive Setup Upon Completing the Workflow. Archive the Study/Studies Archive Dose Report Series · Archive the Series Archive IQ SPECT DC Series To Device: Use MPPS MPPS Setup Setup MPPS Update HIS/RIS Server On Completion Start Another Workflow upon Completion Link Setup Upon Completion, link this Workflow with: Launch As Worklist item Browse Workflow List . Example below: Save All Save None OK Cancel Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 2 of 6 Job Aid | SPECT Dose Report Job Aid Effective Date: 17/01/2020 | HOOD05162003042654 The dose report contains the following information: Really Long Patient Name 123456789 - Radiopharmaceutical Radiation Dose Report X File Edit Help Radioisotope Note : Some parts of this DICOM SR Document cannot be displayed with the current stylesheet . Radionuclide half-life They will not be lost or modified while editing, saving or exporting the DICOM SR Document. Radiopharmaceutical Radiation Dose Report Injected dose Associated Procedure: Radionuclide localization of tumor, Has Intent: Diagnostic Intent Injected date and time Patient Name Really Long Patient Name 123456789 Patient Height Patient ID 19.03.25-13:38:40-DST-1.3.12.2.1107.5.6.1.10054 Patient Weight Date of Birth 3/25/1990 Sex F Effective dose Patient Height Patient Weight Adsorbed dose to organs Referring Physician Content Date 3/27/2019 Content Time 9:50:01 AM Completion Flag| COMPLETE Verification Flag| UNVERIFIED Radiopharmaceutical Administration Person Name Unknown Person Role in Procedure Irradiation Administering Radiopharmaceutical agent Technetium Tc^99m^ mebrofenin Radionuclide ^99m^Technetium Radionuclide Half Life 21624.1 seconds Radiopharmaceutical Administration Event UID 1.3.12.2.1107.5.6.1.10054. Radiopharmaceutical Start DateTime 3/25/2019 1:38:40 PM Administered activity 222 MBq Template="DoseReportNM" Study description="" Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 3 of 6 Job Aid | SPECT Dose Report Job Aid Effective Date: 17/01/2020 | HOOD05162003042654 To view the report: 1. From the syngo Patient Browser, select a dose report labeled as a SPECT Dose Report. 2. Double click the Dose Report icon in the upper right corner. A dose report is opened in the SR Viewer. 3. Scroll down to view the detailed dose information. Patient Browser Patient Applications Transfer Edit View Elter Evaluation Sort Options Help 19.03.20-12:54:43-DST Local Database Test_12 [1000] SPECT DOSE REPORT Scheduler [1000] RecycleBir [1000] March_25_Test DVD name Test_123 tion Date of birth 3/25/1993 Study date 3/25/2019 Patient ID 19.03.25-12:19:19-DST-1.3.12.2.1107.5.6.1.10054 Modality SR 3/251An19 Current Fitter, Off The following radiopharmaceuticals (isotope/pharmaceutical combination) yield a dose report. Isotope selected in MI Apps Pharmaceutical selected in MI Apps 67 Gallium Citrate 111 Indium Labeled WBC 111 Indium HIG 111 Indium Octreotide 123 Iodine NaI (Blocked) 123 Iodine NaI 131 Iodine NaI (Blocked) 131 Iodine NaI 123 Iodine Fatty Acids Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 4 of 6 Job Aid | SPECT Dose Report Job Aid Effective Date: 17/01/2020 | HOOD05162003042654 Isotope selected in MI Apps Pharmaceutical selected in MI Apps 123 Iodine Ioflupane (DaTscan) 99m Technetium Apcitide 99m Technetium Large Colloid 99m Technetium Sulfur Colloid (Solid) 99m Technetium Small Colloid 99m Technetium DMSA 99m Technetium ECD 99m Technetium HMPAO 99m Technetium Disofenin 99m Technetium Mebrofenin 99m Technetium MMA 99m Technetium MAG3 99m Technetium HDP 99m Technetium Sestamibi (Rest) 99m Technetium Sestamibi (Stress) 99m Technetium MDP 99m Technetium HMDP 99m Technetium DTPA 99m Technetium Pertechnetate 99m Technetium PYP 99m Technetium DMSA 99m Technetium Tagged Red Cells 99m Technetium Tetrofosmin (Rest) 99m Technetium Tetrofosmin (Stress) 99m Technetium Technegas 99m Technetium Labeled WBC 201 Thallium Ion 133 Xenon Aerosol 99m Technetium Furifosmin (Rest) 99m Technetium Furifosmin (Stress) 123 Iodine MIBG 131 Iodine MIBG Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 5 of 6 Job Aid | SPECT Dose Report Job Aid Effective Date: 17/01/2020 | HOOD05162003042654 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com Unrestricted Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 Page 6 of 6
- SPECT
- Dose