PEPconnect

Deleting a Calibrated TestPak Lot Video

This video provides instructions on how to delete a calibrated test pack lot when it is no longer in use on your Stratus® CS Analyzer, including how to display all lots that have been calibrated for a method.

A calibrated test pack lot number can be deleted when it's no longer in use from any mode screen. Press F4. The menu option. In this scenario, we're starting from the patient mode screen. From the menu screen, press F3 reports. From the report screen, press F3 calibration status. Press F1 status to display the lots calibrated for the method. Highlight the lot that you want to remove from the system or select next method to navigate to the appropriate test. Press F1 to delete the lot. To return to the main menu, press the paper Advanced Key and the exit key at the same time.

C : 2698 920

  • video
  • delete TestPak lot
  • delete calibrated TestPak lot
  • calibration