syngo.via VA30 – Parallel CPR Ranges
syngo.via VA30 – Parallel CPR Ranges
syngo.via® Parallel CPR Ranges To document condition of vessels, you can create a series of images along the Curved Planar Reformation (CPR). Parallel CPR ranges save a stack of images containing cross- sectional images along the corresponding vessel path. One curved MPR image provides the reference location for the corresponding cross-sectional images. 1. From the upper left hand corner in-segment tools of a CPR segment, select Parallel CPR Ranges. Radial CPR Ranges Parallel CPR Ranges Zoom/Pan Windowing Home Position Rotate along Centerline Show Centerline 2. The Parallel CPR Ranges window opens. Parallel CPR Ranges ? x Distance: 6 mm Image comment Range series name Runoff r* CrossSections MPR Start Cancel This window provides the following parameters: Distance - Defines the distance between pairs of images of the series in mm. Image Comment - Defines a text that is written into the DICOM image comment Ranges Series Name - Defines the name of the result series. The default name uses the name of the selected centerline followed by “CrossSections” and the display type that is currently applied to the cross-sectional segments SIEMENS www.usa.siemens.com/healthcare 641 Effective date 9/8/2016 Page 1 of 2 3. Modify the values to meet your needs and click the Start button. 4. A new series containing the parallel CPR range is created, based on the currently selected centerline. The new series can be found in the Series Navigator and the Findings Navigator. SIEMENS www.usa.siemens.com/healthcare Page 2 of 2
- syngo.via VA30 – Parallel CPR Ranges