syngo.via VA30 - Radial CPR Ranges
syngo.via VA30 - Radial CPR Ranges
syngo.via® Radial CPR Ranges To document condition of vessels, you can create a series of images along the Curved Planar Reformation (CPR). Radial CPR ranges save a stack of images containing multiple CPR views of the displayed vessel from different viewing direction around the vessel. 1. From the upper left hand corner in-segment tools of a CPR segment, select Radial CPR Ranges. Radial CPR Ranges Parallel CPR Ranges Zoom/Pan Windowing * Home Position Rotate along Centerline Show Centerline 2. The Radial CPR Ranges window opens. Radial CPR Ranges ? X Degree steps: 60 Degree Image comment Range series name Runoff r* Curved Range Radial MPR_CURVED Start Cancel This window provides the following parameters: Degree Steps - Defines the angle between pairs of images. The value also defines the total number of images (360° ÷ [degree steps]) Image Comment – Defines a text that is written into the DICOM image comment Ranges Series Name – Defines the name of the result series. The default name uses the name of the selected centerline followed by “Curved Range Radial” and the display type that is currently applied to the CPR SIEMENS www.usa.siemens.com/healthcare 645 Effective date 9/8/2016 Page 1 of 2 3. Modify the values to meet your needs and click the Start button. 4. A new series containing the radial CPR range is created, based on the currently selected centerline. The new series can be found in the Series Navigator and the Findings Navigator. SIEMENS www.usa.siemens.com/healthcare Page 2 of 2